110-392
27
Troubleshooting
Many problems can be resolved by restoring the factory defaults. If this solves the problem, be careful when
restoring your operational settings in order not to repeat the problem.
To restore the factory default settings:
1. Place the Power Switch / Circuit Breaker on the rear of the unit in the OFF position.
2. Press and hold the Power Button on the front panel while returning the Power Switch / Circuit Breaker to the
ON position.
WARNING:
Service should only be performed by fully qualified personnel. Extreme caution is required as
hazards are present when servicing this equipment.
Problem
Possible Causes
Corrective Action
Unit does not run
(digital displays blank)
No power to unit
Check that the electrical cord or wiring is secure and
connected to an operating electrical source.
Check that Power Switch / Circuit Breaker on rear of unit
is ON.
Unit does not run
(three decimal points appear
on Temperature Display, two
decimal points on
Pressure/Flow Rate Display)
Unit in Standby mode
Press Power Button on front panel.
No fluid circulation
Insufficient fluid in reservoir
Blockage in circulating system
Pump is not operating
Add fluid to reservoir.
Remove blockage.
Replace pump.
Insufficient circulation
Fluid viscosity too high
External tubing diameter too
small
Restrictions in fluid lines
Low line voltage
Replace with lower viscosity fluid.
Replace with larger diameter tubing.
Check and correct as required.
Check and correct as required.
Unit does not cool or cooling
is insufficient
Dust build up on air filter or
condenser
Blocked air ventilation screens
Excessive heat load
Ambient air temperature too
high
Low or high line voltage
Faulty temperature sensor
Clean air filter and/or condenser as required.
Remove blockages as required.
Check that heat load does not exceed capacity of Chiller;
correct as required.
Decrease ambient air temperature.
Check and correct as required.
Check the compressor upper, evaporator inlet and
evaporator outlet temperature sensor readings (see
“Diagnostic Mode”). If any of these temperature readings
is -50°C, the sensor needs to be replaced.
Unit does not heat
OTP has tripped due to
Insufficient fluid in reservoir
Add fluid to reservoir and reset OTP.
Fault code 10 on display
Extreme electrical line
interference
Triac failure
Plug unit into another power source. If problem persists,
triac has failed.
Replace triac or triac driver as required.
Содержание DuraChill 58 Series
Страница 1: ...User Guide DuraChill Air and Water Cooled 1 5 HP Chillers 110 392 27 June 2016...
Страница 2: ......
Страница 32: ...110 392 30 Electrical Specifications...
Страница 39: ...110 392 37 Appendix Flow Schematic All Models...
Страница 40: ...110 392 38 Electrical Diagram 208 230V 1PH 50 60Hz Models without heater option...
Страница 41: ...110 392 39 Electrical Diagram 208 230V 1PH 50 60Hz Models with heater option...
Страница 42: ...110 392 40 Electrical Diagram 208 230V and 380 460V 3PH 50 60Hz Models without heater option...
Страница 43: ...110 392 41 Electrical Diagram 208 230V and 380 460V 3PH 50 60Hz Models with heater option...
Страница 45: ...110 392 43 DI Control Wiring H2 H3 Option Wiring H2 24 VDC H3 12 VDC...