2.

 

OPERATION

 

2.1. Turning the instrument ON/OFF 

To turn the instrument on, press and hold down the MODE/SET button. If 

the  battery  is good,  the audio and vibration alarms are initiated for a 

short period of time, and then the display  briefly  shows  the  firmware 

version  and  battery charge level before  initiating  self-test procedures. 

The message “TEST” is displayed during this operation.  
To turn the instrument OFF, go to the SETTINGS menu and select POWER 

OFF. The LCD will display the “POWER OFF” message and after a couple 

of seconds goes blank. Now the instrument is turned off.  

2.2. Going through self-tests and background calibration 

After the self-tests are completed the audio and/or vibration alarms are 

activated for approximately 1 second before the instrument initiates the 

background calibration. The message "CALIBRATION" is displayed on the 

LCD during this process, in which the instrument measures the gamma 

radiation  background  and calculates alarm levels  given  a  value of n 

coefficient2. After background calibration is finished, PM1704A is ready 

for normal operation.  

2.3. Normal operation (Search mode) 

When self-tests and background calibration are finished the instrument 

is ready for operations and automatically  goes  into  Search mode. In 

Search mode, the instrument displays the current value of gamma and 

neutron  radiation  (for gamma-neutron models only) in  counts per 

second (cps) and  automatically compares it  with the threshold  value 

obtained during background calibration. 
This allows the user to effectively search and locate radioactive sources. 
When a radioactive source is detected (the calculated  threshold is 

exceeded), the instrument’s  visual,  audible  and/or  vibration  alarms  are 

triggered. As the PM1704A  moves closer to the  radiation source, the 

frequency of audio and/or vibration signals increases thus enabling the 

user to determine the direction of the source location.  
When the alarm signals reach maximum frequency it is recommended to 

run  the  background calibration procedure (see section 2.3.1.)  while 

keeping the

 location of the instrument unchanged

. After background 

calibration is completed, the user may continue to seek the location of 

the  radiation source. If necessary, this procedure may be repeated 

several times until a gamma radiation source is located. 

                                                                    

2

 See section 2.3.2 for more information. 

Содержание PM1704A

Страница 1: ...2013 2015 Polimaster Ltd All Rights Reserved SPECTROSCOPIC PERSONAL RADIATION DETECTORS PM1704A models PM1704A PM1704A M PM1704A GN PM1704A GNM OPERATIONS MANUAL http www polimaster com...

Страница 2: ...nt 13 2 3 3 False alarms 14 2 3 4 Autocalibration 15 2 4 Measurement mode 15 2 4 1 Setting new thresholds 16 2 4 2 Resetting statistics restarting measurement 17 2 5 Acquiring gamma scintillation spec...

Страница 3: ...ors SPRDs PM1704A are designed to detect search locate and identify gamma emitting radioactive sources as well as detect neutron radioactive materials gamma neutron models and measure dose and dose ra...

Страница 4: ...strument in water 2 Do not store the instrument with the battery installed The battery may leak once depleted causing severe damage to the instrument and voiding the warranty 3 Keep the instrument at...

Страница 5: ...urces There are four different modifications of PM1704A two gamma only and two gamma neutron They differ in the number and kind of detectors used Gamma only Spectroscopic Personal Radiation Detectors...

Страница 6: ...N A from 0 01 to 999 cps Indication range gamma 0 01 130 Sv h 1 R h 13 mR h 0 01 Sv h 13 Sv h 1 R h 1300R h 0 01 130 Sv h 1 R h 13 mR h 0 01 Sv h 13 Sv h 1 R h 1300R h Accuracy of exposure rate measur...

Страница 7: ...ta points 100 spectra files Environmental temperature range humidity 4 to 122 F 20 to 50 C up to 98 at 104 F 40 C Water tightness IP65 Drop test 1 5 m Dimensions 75 87 38 mm 102 87 38 mm 102 87 38 mm...

Страница 8: ...bols external power source audio and vibration alarms are disabled audio alarm is enabled vibration alarm is enabled Bluetooth module is enabled GPS module is enabled battery charge level Search mode...

Страница 9: ...on removal The instrument except the basic PM1704A model has a clip and may be fasten to the user s belt The clip may be removed using a screwdriver or similar tool and easily replaced later as shown...

Страница 10: ...ery compartment cover see picture below Then insert a single AA type battery into the battery holder following polarity instructions on the label reinstall the battery compartment cover and tighten th...

Страница 11: ...echargeable NiMh batteries Therefore after the first time the instrument is turned on the user has to select the proper battery type in the instrument SETTINGS menu as shown below If a rechargeable ba...

Страница 12: ...mal operation 2 3 Normal operation Search mode When self tests and background calibration are finished the instrument is ready for operations and automatically goes into Search mode In Search mode the...

Страница 13: ...new background and recalculate thresholds To do this press and hold the SELECT button until the Search mode menu appears on the display and select CALIBRATION The unit is now acquiring a new backgroun...

Страница 14: ...value of the n coefficient is the more sensitive the instrument is to changes in background radiation however this also increases the rate of false alarms The larger the value of the n coefficient is...

Страница 15: ...ng the neutron detector In the event that the threshold value is exceeded audible and or vibration alarms are activated to warn the user of the presence of a gamma neutron radiation source nearby The...

Страница 16: ...values press and hold the SELECT button until the Measurement mode menu appears on the display then press the MODE SET button to select THRESHOLD 1 or THRESHOLD 2 to change the corresponding threshold...

Страница 17: ...ra PM1704A is designed to allow the user to acquire gamma scintillation spectra and perform express radioisotope identification This is a critical feature that helps to distinguish between real threat...

Страница 18: ...nformation SPEC TIME accumulation time in seconds COUNT RATE OK LOW HI message count rate analog scale Please try to keep such a distance from the source and the instrument that during all time of spe...

Страница 19: ...T button to see details The following information will be displayed NUKLID name of the radioisotope e g EU152 TYPE type of the radioisotope e g IND CONF the confidence level e g 99 PEAKS number of pea...

Страница 20: ...cally from Search mode after the given time interval of the instrument s alarm If this value is set to DISABLED then it will not run the identification process automatically 2 6 2 Displaying peaks Inf...

Страница 21: ...ange DE thresholds or reset accumulated DE value it is needed to enter the DE measurement mode menu For this purpose press and hold the SELECT button until the DE measurement mode menu appears on the...

Страница 22: ...he GPS module Setting the instrument date and time Configuring history settings Configuring display settings Configuring audible alarm volume Selecting battery type Viewing information about the instr...

Страница 23: ...lt in user software and manual PM1704 DATA contains all stored spectra and history data PM1704 INFO contains service information To start PM1704A Built in Software go to the new PM1704 SOFT drive and...

Страница 24: ...ocedure must be performed at the ambient temperature not exceeding 85 F 30 C Isotope identification library shell be set to ANSI see sec 2 6 1 of Operation Manual Please follow guidelines described in...

Страница 25: ...t the manufacturer or authorized technical center in case of any questions NOTE Do not perform this operation in a preventive manner in case the Energy Calibration Check is passed correctly To perform...

Страница 26: ...efficients Please make sure that difference between current and new coefficients value is less than 30 in the case shown above the difference is calculated as 124 2 126 6 2 4 which is less then 30 the...

Страница 27: ...ced to low battery power or poor battery quality Use only fresh high quality alkaline batteries The PM1704A is an extremely low power device and as such the unit should be allowed to stand without the...