17
indication, audio and/or vibration alarms with one
second interval when Safety Alarm threshold is
exceeded.
The manufacturer’s preset value of Safety Alarm
threshold is 10
µ
Sv/h (1000
µ
R/h). If necessary the
Safety Alarm threshold can be set using
PM1703GNA-II and PM1703MA-II User Software.
3. SETUP
Using the detector front panel buttons it is possible
to adjust several parameters of the instrument. The
number and combination of settings enabled for
editing in your particular device depends on factory
pre-sets and may vary in each instrument. See
below the complete list of settings that may be
enabled for editing in the instrument:
•
n coefficient, gamma may be setup in the
range of 1.0 to 9.9 with increment of 0.1. The
manufacturer’s recommended value is 5.3;
•
n coefficient, neutron may be setup in the
range of 1.0 to 9.9 with increment of 0.1. The
manufacturer’s recommended value is 5.0;
•
Enable or disable the Sound alarm on
reaching the alarm threshold;
•
Enable or disable the Vibration alarm on
reaching the alarm threshold.
To enter the Setup Menu, press and hold down the
Mode button
. The “CAL” message will be
displayed first. Continue to hold the Mode
button down until the screen displays the first
setting that can be adjusted.
NOTE
: Depending on the options enabled in each
particular instrument, the set of properties available
for adjustment can differ in two devices of the same
model. Furthermore, some detectors may not enter the
Setup Menu at all, since the adjustment of all settings
can also be disabled using the User Software.