1.3 Specifications
1
Operating mode:
- indication of photon radiation DER;
- indication of photon radiation DE;
- indication of the instrument’s number;
- PC data exchange;
-
search;
- indication of the audible alarm;
- indication of the vibration alarm;
-
set
mode;
- indication of partial and critical battery
discharge;
- audible, vibration and visual (blinking red
LED) alarming at exceeding of the preset DE
or DER thresholds.
2
DER measurement range:
- for the
РМ
1621M modification
0.1
µ
Sv/h - 100 mSv/h;
- for the
РМ
1621M
А
modification
0.01
µ
Sv/h – 2.00 Sv/h.
- DER indication sub-range
0.01 – 9.99
µ
Sv/h;
10.0 – 99.9
µ
Sv/h;
100 - 999
µ
Sv/h;
1.00 – 9.99 mSv/h;
10.0 – 99.9 mSv/h;
100 - 200 mSv/h;
100 - 999 mSv/h (
РМ
1621M
А
);
1.00 – 2.00 Sv/h (
РМ
1621M
А
).
- DER analogue scale (seven
segments) in a logarithmic gauge
Number
of
indicated
segments (from left to
right) corresponds to the following DER
threshold values on the LCD display:
- one segment -
≥
0.1
µ
Sv/h;
- two segments -
≥
1
µ
Sv/h;
- three segments -
≥
10
µ
Sv/h;
- four segments -
≥
100
µ
Sv/h;
- five segments -
≥
1 mSv/h;
- six segments -
≥
10 mSv/h;
- seven segments -
≥
100 mSv/h.
3
Maximum permissible intrinsic relative
error of DER measurement
±
(15 +
К
1
/
!
+
К
2
!
) %,
where
!
- DER value, mSv/h;
К
1
– coefficient 0.0015 (mSv/h);
К
2
– coefficient 0.01 (mSv/h)
-1
.
4
DE measurement range
1.0
µ
Sv - 9.99 Sv;
Maximum
permissible
intrinsic relative
error of DE measurement
±
15 %.
5
DE indication range
0.01
µ
Sv – 9.99 Sv.
Indication sub-ranges:
0.01 – 9.99
µ
Sv;
10.0 – 99.9
µ
Sv;
100 – 999
µ
Sv;
1.00 – 9.99 mSv;
10.0 – 99.9 mSv;
100 - 999 mSv;
1.0 – 9.99 Sv.
7
Содержание PM1621M
Страница 1: ...X RAY AND GAMMA RADIATION PERSONAL DOSIMETER РМ1621M PM1621MA OPERATING MANUAL 2 ...
Страница 13: ...Figure 1 2 General overview of the instrument 14 ...
Страница 14: ...Figure 1 3 The PM1621M with a removable clip 15 ...
Страница 21: ...Figure 2 1 Choice of operating mode indication of the instrument 22 ...
Страница 22: ...Figure 2 2 Setting the DE threshold values 23 ...
Страница 24: ...Figure 2 4 Audio control mode control of the sound volume 25 ...