8
2) to download the following information from the
dosimeter to computer:
- dosimeter ID (serial number);
- DER and DE alarm threshold values;
- DER and DE data logging interval;
- time delay to record the first event of DER and
DE history;
- DER and DE history (date, time, event, value)
according to preset data logging interval;
- DER and DE history (date, time, event, value) of
user-initiated recorded history;
- DER and DE history (date, time, event, value) at
alarm events;
3) to set the following parameters in the dosimeter
from the computer:
- current computer time and date;
- DER and DE alarm threshold values;
- DER and DE data logging intervals;
- time delay to record the first event of DER and
DE history;
- reset accumulated DE and DE accumulation time;
4) reading from the dosimeter into PC: dosimeter
number, current DE and DER values by the
ISO15693 interface.
1.3.20
Setting time
60 s
1.3.21
Power supply
- battery 1.5 V, AAА (LR03), no less than
1100 mA/h;
- external power supply (computer USB port).
1.3.22
Battery lifetime (full charge, 1100 mAh) in
normal operation (average DER not more than
0.3 µSv/h, backlight, sound, light and vibration alarms
activated for not more than 20 s/24 hours)
no less than 20 days
1.3.23
Ingress protection
IP65
1.3.24
Operation conditions:
- air temperature
- relative humidity
- atmospheric pressure
-20 to + 50 °С;
98 % at 35 °С;
84 to 106.7 kPa.
1.3.25
Instrument is resistant to:
- sinusoidal vibration in frequency range 5 –
35 Hz and bias amplitude for frequencies lower
than transition frequency 0.75 mm;
- shocks with peak acceleration 100 m/s
2
, duration
of shock pulse 2-50 s, shock rate 60 - 180
shocks/min
1.3.26
Drop test
1.5 m onto hard wooden surface.
1.3.27
Dosimeter withstands dc and ac magnetic fields 800 A/m
1.3.28
Dosimeter withstands electrostatic discharges, IDT (air discharge 8 kV, contact discharge 6 kV)
1.3.29
Dosimeter corresponds to ANSI 42.34, section 9.11 requirements by the level of emitted radio-noise
1.3.30
Dosimeter is proof against the action of radio frequency electromagnetic fields up to 30 V/m (rigidity
degree 4) in the frequency range 80 – 1000 MHz and in the frequency range 800 – 960 MHz and 1.4 – 2.5 GHz
(under conditions of noise emission from digital radio telephones), performance criterion A
1.3.31
Weight, max
Weight in package, max
0.09 kg;
0.4 kg.
1.3.32
Overall dimensions, max
71 х 59 х 20 mm
1.3.33
Reliability:
- mean time between failures (MTBF), min
- average service life, min
- mean time to recovery (MTTR), max
20000 h;
10 years;
60 min.
N o t e – For more information please contact the manufacturer or visit www.polimaster.com.