Instructions for Use 2019-03-20
The initial settings are shown below. These settings can be changed.
Primary Functions
:
Speed, Rotation Mode
1. Select a memory number for the standby display by pressing the
Plus or Minus Switch.
2. Press the Select Switch to choose one of the primary functions.
3. Press the Plus or Minus Switch to change the setting.
* The display will go back to the standby display if 5 seconds (initial
setting) elapses without a switch being pressed.
Speed Settings:
< OTR Mode >
100, 300, and 500 rpm
< Normal Mode >
50, 100, 150, 200, 250, 300, 400, 500, 600, 800, and 1000
rpm
Torque Settings:
< OTR Mode >
0.2, 0.4, 0.6, 0.8, and 1.0 Ncm
< Normal Mode >
0.2, 0.4, 0.6, 0.8, 1.0, 1.5, 2.0, 2.5, 3.0, 4.0, and 5.0 Ncm
This function can also be turned off: TRL (torque reverse-less).
Rotation Mode:
Normal (Fwd: Forward, Rev: Reverse) or OTR
Standby Display
Rotation Mode
Speed
Torque
**
These functions are available only when connected to the Endostar Navigator.
• If the torque limit is too high, the file could jam inside the canal and break.
• The torque settings must be changed depending on the root canal condition and the file.
• If the torque reverse seems to be activated too frequently increase its value.
• When OTR mode (motor runs back and forth continuously) seems to be triggered too frequently, or it is triggered immediately after starting the normal rotation, increase the
torque setting by one line.
Setting
Memory
M1
M2
M3
M4
M5
M6
Speed (rpm)
500
300
500
500
300
500
Reverse Torque (Ncm)
0.2
0.2
0.2
1.0
1.0
1.0
Rotation Mode
OTR
OTR
OTR
Normal
(Forward)
Normal
(Forward)
Normal
(Forward)
Torque Slow Down
—
—
—
Off
Off
Off
Linked Function
**
On
On
Off
On
On
Off
Apical Reverse or Stop
**
Reverse
Reverse
—
Reverse
Reverse
—
Auto Start or Stop
**
On
On
Off
On
On
Off
Apical Slow Down
**
—
—
—
Off
Off
Off
Apical Torque Reduction
**
—
—
—
Off
Off
Off
Torque reverse values may differ somewhat depending on the motor and
contra angle.