background image

3

g)  Refer to section “h” in previous section for fitting the vest, head cap, wrap,

blanket, seat, or bladder.

h) Plug in the Chiller power cord and turn the Chiller on. See Chiller label for proper

voltage.

i)

Connect the transformer power supply into the cooling reservoir and then plug
into a wall outlet. See power supply label for proper voltage.

j)

Flip the cooling reservoir power switches to the “On” position.

k) Water will be circulating through the Chiller and the garment as follows:

• Water will be circulating from the cooling reservoir to the chiller and back to the

cooling reservoir. You can see this flow by slightly raising the cooling reservoir lid.

• Water will be circulating from the cooling reservoir to the vest, head cap, wrap,

blanket, seat, or bladder and back to the cooling reservoir. Slightly raise the
cooling reservoir lid and ensure there is water flowing through the top return flow
tube to the cooling reservoir from the vest, head cap, wrap, blanket, seat, or
bladder.

l)

After all components have been filled and run with water, check the reservoir to
assure the water level is well above the pump. Add additional water if necessary.

m) Set the Chiller temperature: Hold down the “Set” button for 5 seconds. The Chiller will

beep and the display screen will flash. Once it does, use the “up” and “down” arrows
to select the desired temperature between 40 degrees F and 70 degrees F. Then
press the set button again. The screen will switch back to display the current water
temperature. To help the cooling process, add ice to the water in the cooling
reservoir. 

Note: Please read the instructions in this manual and included Chiller

instructions for Care & Maintenance, Troubleshooting, Operating Instructions,
Cautions and Warnings.

Ending Session

a) Unplug all components from the power source and disconnect the vest, head cap,

wrap, blanket, seat, or bladder and tubing.

b) Disconnect the Chiller from the cooling reservoir if a chiller system was

purchased. Follow Chiller instructions.

c) To empty the cooling reservoir use

the Drain Accessory which is located on
a clip on the inside of the cooling
reservoir lid. Connect the drain accessory to the female coupling on the end of
the tubing coming from the cooling reservoir, placing the open end of the tubing
in a sink or area where water may be drained without incident/injury. Run the
system until the rate of draining water slows significantly. Note: Drain Accessory
will not drain all water from the system. If pump is left running too long without
water circulation, the motor may burn out.

d)  Flip the power switch to the “Off” position and pour out the remaining 1-2 inches

of water. Empty the reservoir and wipe out after each use .  

e)

Tabbed couplings

have a metal or plastic tab on the female side, depress this tab

and gently pull the couplings apart. 

Breakaway couplings

are simply pulled apart.

Drain Accessory

Содержание Cool Flow

Страница 1: ...Cool Flow CoolOR PERSONAL COOLING SYSTEMS I N S T R U C T I O N S POLAR PRODUCTS 800 763 8423 polarproducts com...

Страница 2: ...using and operating the Polar Cool Flow or CoolOR cooling system Note Each Polar Cool Flow or CoolOR system has been fully tested in the factory with a distilled water and alcohol solution There will...

Страница 3: ...ocked together Breakaway couplings are simply pushed together e The temperature ow control valve will be located on the cooling reservoir or with a separate attachment depending on your system If it i...

Страница 4: ...ollow all Instructions Cautions and Warnings on this page the Cool Flow and CoolOR instructions in this manual the chiller instructions and the instructions on the product s before operating this syst...

Страница 5: ...mperature between 40 degrees F and 70 degrees F Then press the set button again The screen will switch back to display the current water temperature To help the cooling process add ice to the water in...

Страница 6: ...p wrap blanket seat or bladder is always stored in a clean dry environment with good ventilation Care of system Once every two weeks and prior to extended storage pour 16 ounces of Isopropyl Alcohol r...

Страница 7: ...r level in the cooling reservoir Con rm that the pump is completely submerged in the water Check for kinks Ensure there are no kinks in the hoses insulated tubing or vest head cap wrap blanket seat or...

Страница 8: ...h ice Note the line inside the cooling reservoir Ensure that the temperature control knob is turned to maximum cold COLDEST If the Chiller System has been purchased check the chiller for proper operat...

Страница 9: ...ounter a problem with your Cool Flow or CoolOR System please call Polar s Customer Service Department at 1 800 763 8423 to obtain a Returned Goods Authorization number RGA To obtain warranty service o...

Страница 10: ...mbroider your name on any of our vests Call Polar for details Cooling reservoir size and number of systems 13 quart one 1 or two 2 person system 15 quart one 1 or two 2 person system 19 quart one 1 th...

Страница 11: ...ry quick disconnect female coupling with plastic tab See coupling pictures on page 8 for reference Available accessories and replacement parts Pumps tube insulation and all other components are availa...

Страница 12: ...800 763 8423 polarproducts com POLAR PRODUCTS 3380 Cavalier Trail Stow OH 44224...