FCC ID:
M5ZANA
Example:
C1
283FA60B000000EC000F8AD0AB<CR>
This field is the mode indicator, the LS-bit which indicates the
service state of the transmitter, C1=normal, C0=service mode.
C1
283FA60B000000EC
000F8AD0AB <CR>
This
field
is
the
unique
serial number of the 1-Wire sensor.
C1283FA60B000000EC
000F
8AD0AB <CR>
This is analog data field stored MSByte first as a 12-bit number in
a 16-bit field. The number shown is 000F (15) or 15/4096 of full
scale. If this is a 0-20 milliamp input this would be .07 MilliAmp.
If the input is 0-5 Volts this would be 0.0183 Volts.
C1283FA60B000000EC000F
8AD0
AB <CR>
This field is the CRC-16 error check as was originally received and
checked. This CRC is over the first 11 bytes of the packet starting
with the mode flags and ending with inclusion of the temperature
data.
C1283FA60B000000EC000F8AD0
AB
<CR>
This field is the mod 256 sum of all the ASCII character values in
the response but does not include the <CR>.
C828AE9105000000AA2D2DD474AB
<CR>
This is the CR terminator, 0Dhex.