1.8. WATER FLOW DIAGRAM
1.9. SHOOTING TROUBLE
Problem
Action
The power cord is connected to the mains,
the electrical switch S is ON, and the red
light on the switch does not turn on.
Measure and check that the mains voltage is 230 V.
Check the status of the switch.
The water still is on, water is coming out of
drain D and the green L1 light is off.
Follow this order:
Remove the thermostat protector T and press inwards.
Check level switches Ne and Nd.
Check contactor C1.
Check heaters R1 and R2.
The water still is on and the water tank is
not at its maximum capacity, no water
circulates.
Replace the electrovalve E.
If the water level in the boiler is below the
temperature sensor and the sensor triggers
Press the button REARME. (see fig.2.)
Note: It will not be allowed to repair any water still that is not properly cleaned.