Page 8/44
Van ’t Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
F. 31 (0) 10 521 19 41
E-mail: [email protected]
Website: www.tamson.com
VAT: NL 80 66 34 984 B01
Bank account no.: Rabobank 16 01 00 046
Chamber of commerce 27 16 95 41
IBAN Code: NL95 RABO 0160100046
Tamson Instruments bv
Rev. 2.1 UK 0517
Airflow
Air must pass the apparatus freely. The air transports
the heat from inside the apparatus to the environment.
When airflow is blocked the apparatus will not perform
to its specifications and it may harm the equipment or
even cause mechanical damage.
Dust
Do not place the bath in a dusty environment. Dust will
block the condenser unit inside the apparatus which will
lead to severe mechanical damage.
Regularly check the apparatus and condenser unit
for dust. Remove dust with a vacuum cleaner. Do
NOT remove dust with compressed air.
Leaking
Inside the bath the cold parts condensate and ice can
be build. When the apparatus is switched-off this melt
water may leak onto the floor. Do not place the
apparatus on a floor which can be damaged by this melt
water. The amount of water is limited to a few mL/hr.
4.5
Initial use
Due to transportation allow the bath a 24 hour
period for stabilization. This applies also when the
apparatus has been tilted or fallen.
Do not switch on power, because lubrication-oil inside
the compressor system has to run into the capillary.
It will take several hours before the oil has flown back
into the compressor unit. When the apparatus is
immediately turned on after it has been tilted, severe
damage may occur to the compressor unit due to
insufficient lubrication.
AIR FLOW