RespiraSense
™
Instructions For Use
RespiraSense
™
PDS-801-007 Revision 2
Page
34
of
77
CAUTION
During any step of setup warning or error messages may be displayed. Read all
messages carefully and take action when prompted.
Step 1: Registration of a Lobe
In this Step, a fully charged Lobe is registered to a specific patient. This uniquely pairs the Lobe and
patient and allows retrieval of patient data at later steps.
1.
Retrieve fully charged Lobe from Charging Station. A fully charged Lobe displays a solid Green
LED when in the Charging Station.
2.
Wipe down Lobe in accordance with hospital, administration, and/or local government laws.
3.
Launch the RS App and select
Add New Patient
on the Dashboard.
4.
A Lobe can be renamed when registering the lobe once the lobe is selected the patient MRN
can be entered.
See Figure 8.
A patient can be registered to a different lobe for example if one
lobe’s battery dies once the patient MRN name is identical the patient’s information will be
transferred to the new lobe and be seen on the RS app.
5.
Verify that alarm limits, Lobe identifier, averaging window and patient ID are correct on
confirmation screen.
CAUTION
Verify that the upper and lower respiratory rate thresholds are in accordance with
current hospital, administrative, and/or local government law limits.
6.
Select
Yes
if details are correct
–
this will complete the renaming process.
7.
Select
No
if incorrect
–
this will cancel the renaming process.
8.
Confirm rename is successful. On successful rename the following occurs:
●
Confirmation successful.
●
Lobe beeps twice.
●
Lobe LED is set to Flashing White.
9.
The RS device can now be assembled and applied to the patient in accordance with
PDS-801-
008: RespiraSense Setup and Device Instructions for Use
.
Summary of the registration process.
Solid Green LED
(When in Charging
Station)
Register New
Patient
Select from List
Add MAC or scan
QR code on any
accessories
Device Confirmation -
Flashing White LED