Changing Gas Cylinders 

When either oxygen or nitrous oxide gas supply pressure drops to approximately 35 psi (2.41 
bar), a whistle alarm will sound to alert the user. Check the line pressure gauges to confirm 
which gas is low and then replace that cylinder immediately following these steps:

1. Use the cylinder wrench to turn the yoke valve clockwise to firmly close the low cylinder.

(You may want to close the opposite cylinder while performing although it is not necessary.)

2. Turn the regulator T-bar counter-clockwise to loosen the regulator from the cylinder. Lift
the regulator  off the yoke and replace  the cylinder.  Ensure  the regulator  still has the nylon
washer around the gas inlet. There is no need to undo or loosen any other connections.

3. Remove any protective wrap from around the yoke of the new cylinder. Remove any FULL
tags from new cylinder once you begin using it.

4. Place the regulator over the yoke of the new cylinder, lining up the index pins and gas inlet
so they slide into their receiving holes.

5. Hand-tighten the regulator T-bar (clockwise, to the right) so the regulator is firmly attached
to the cylinder yoke.

PLYMOUTH MEDICAL

135 PLYMOUTH ST, SUITE 410

 

BROOKLYN, NY

 

11201

TEL – 888.392.5076

FAX – 888.392.1455

WEB – 

PLYMOUTHMEDICAL.COM

COPYRIGHT © 2020 PLYMOUTH MEDICAL LLC. ALL RIGHTS RESERVED.

V3-01-22

Nitronox Plus®

Quick Start Guide

Содержание Nitronox Plus

Страница 1: ... PRESSURE GAUGE IS NOT WITHIN THE 30 40 PSI GREEN ZONE 4 Place a patient circuit onto the front inlet and connect mouthpiece or mask to blue tubing Ensure the magenta tubing is pointed away from patient provider s 5 Turn the N2O Key Lock to the ON position to start administering the gas mixture 6 Set the desired concentration of nitrous oxide by adjusting the N2O dial on the front of the device It...

Страница 2: ... System for the first time Take the hose with the mouthpiece When you feel pain anxiety starting or when instructed hold the mouthpiece in your mouth and take slow deep breaths You will hear the pain relief gas being delivered Keep the mouthpiece in your mouth as you breathe in and out Within a few breaths you will start to feel more relaxed and the pain will be lessened You may also feel a little...

Страница 3: ... green as there is still gas in the lines no need to bleed the lines 6 Take the patient circuit off the mixer front inlet and dispose of the circuit magenta mouthpiece mask if used 7 Wipe the mixer off with disinfectant sani cloth do not spray product directly on device 8 The system is now shut down which should be done at the end of each day 9 We recommend removing the keys and storing them in a ...

Страница 4: ...ulator off the yoke and replace the cylinder Ensure the regulator still has the nylon washer around the gas inlet There is no need to undo or loosen any other connections 3 Remove any protective wrap from around the yoke of the new cylinder Remove any FULL tags from new cylinder once you begin using it 4 Place the regulator over the yoke of the new cylinder lining up the index pins and gas inlet s...

Страница 5: ...s oxide gas and oxygen for procedures You will go through 3 oxygen tanks to every 1 nitrous tanks per below 11 The device is ready for use ALWAYS LISTEN FOR LEAKS AT THE CONNECTIONS HISSING SOUND IF A LEAK IS DETECTED TIGHTEN THE FITTING TO THE REGULATOR T BAR WHERE THE HOSE SCREWS INTO THE SIDE OF THE TANK DEVICE IF WHISTLE ALARM SOUNDS DISCONTINUE PATIENT USE IMMEDIATELY AND SHUT OFF GAS SUPPLY ...

Отзывы: