Page 22/22
Attention
:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the
“plusoptiX 16”, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result. The
“plusoptiX 16” is not compatible to HF SURGICAL EQUIPMENT.
Recommended safety distances between portable and mobile HF telecommunication devices and the “plusoptix 16”
The “plusoptiX S16” is intended for use in an electromagnetic environment with controlled HF disturbance variables. The
customer or user of the “plusoptiX S16” can contribute to avoiding electromagnetic disturbances, by complying with the minimum
distance between portable and mobile HF telecommunication devices (transmitters) and the “plusoptiX S16” – depending on the
output power of the communication device, as stated below.
Maximum output power of the
transmitter in W
Safety distance depending on the transmission frequency (in meters)
150 kHz - 80MHz
d = 0.35 √ P
80 MHz - 800MHz
d = 0.35 √ P
800 MHz - 2,7GHz
d = 0.7 √ P
0.01
0.035
0.035
0.07
0.1
0.11
0.11
0.22
1
0.35
0.35
0.7
10
1.11
1.11
2.21
100
3.5
3.5
7
For transmitters, whose maximum nominal output is not stated in the above table, the recommended safety distance in meters
(m) can be calculated using the equation from the relevant column, wherein P equals the maximum nominal output of the
transmitter in watts (W) as per information by the transmitter’s manufacturer.
COMMENT 1 At 80 MHz and 800 MHz, the higher frequency range applies.
COMMENT 2 These guidelines may not be applicable in all cases. The distribution of electromagnetic quantities is influenced by
absorption and reflections from buildings, objects and people.