6. Error messages/malfunctions
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed
(Example: Error No. 2).
Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked and if
necessary, the corresponding measures are to be taken:
Error No.
Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses in
fl
uence the mea
surement result.
Reason: hand was moved during the Measurement (Artefact).
ERR 3
The in
fl
ation takes too long. The cuff is not correctly seated.
ERR 5
The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another reading
following di rections carefully. Contact you doctor if you continue
to get unusual readings.
ERR8
Pressure in cuff is over 290mmHg
Malfunction
Remedy
The display remains empty when
the instrument is switched on
although the batteries are in
place.
1.Check batteries for correct polarity and if
necessary insert correctly .
2.If the display is unusual, re -insert batteries or
exchange them.
The device frequently fails to
measure the blood pressure
values, or the values measured
are too low (too high).
1. Check the positioning of the cuff.
2. Measure the blood -pressure aga in in peace and
quiet under observance of the details made under
point 5.
Every measurement produces a
different value although the
instrument functions normally
and the values displayed are
normal
1. Please read the following information and the
points l isted under «Common sources of error».
Repeat the measurement.
Please note: Blood pressure
fl
uctuates
continually so successive measurements will
show some variability
.
12