NOTE
The exposure values will automatically change according to the selected
patient size, image resolution and ULD setting. The preset exposure values
are shown in section "Adjusting exposure values for current exposure" on
11.8 Adjusting exposure values for current exposure
NOTE
Always try to minimize the radiation dose to the patient.
The exposure values have been preset at the factory for each patient size,
image resolution and ULD (Ultra Low Dose) setting. The preset exposure
values are average values and they are only meant to guide the user.
The preset exposure values are shown in the following tables.
Factory presets for voxel sizes 450 or 600 micrometres
PATIENT SIZE
kV VALUE
mAs VALUE
mAs VALUE
WITH ULD
Child (XS)
90
36
7.1
Small adult (S)
90
40
8
Medium-sized adult (M)
100
32
6.3
Large adult (L)
110
28
5.6
Extra large adult (XL)
120
25
5
Factory presets for voxel size 300 micrometres
PATIENT SIZE
kV VALUE
mAs VALUE
mAs VALUE
WITH ULD
Child (XS)
90
45
9
Small adult (S)
90
50
10
Medium-sized adult (M)
100
40
8
Large adult (L)
110
36
7.1
Extra large adult (XL)
120
32
6.3
Factory presets for voxel size 150 micrometres
PATIENT SIZE
kV VALUE
mAs VALUE
mAs VALUE
WITH ULD
Child (XS)
90
71
14
Small adult (S)
90
80
16
Medium-sized adult (M)
100
63
12.5
Large adult (L)
110
56
11
Extra large adult (XL)
120
50
10
11 3D exposure
50 Planmeca Viso G7
User's manual