EXPOSURE VALUES

18  Prostyle Intra X-ray Unit

User's manual

9.2

Preprogrammed settings values

NOTE Two sets of exposure values (time/kV) have been pro-

grammed for each exposure region: one for adult
mode and one for child mode.

NOTE The exposure values are programmed corresponding

the normal mode density value 0 (factory preset va-
lue).

NOTE In the digital imaging mode the exposure time values

are automatically scaled according to the density va-
lue used in digital imaging mode. The factory preset
density value in the digital imaging mode is -4.

These values can be programmed by the user, see sec-
tion 10.2 “Programming the preprogrammed settings”
on page20. Th
e recommended exposure values are gi-
ven in section 12 “EXPOSURE VALUE TABLES” on
page 23.

Changing the density value will change the selected ti-
me value as follows: one density step equals to one ti-
me step. The negative density value shortens the selec-
ted time value, whereas the positive value lengthens it.

NOTE These values are for speed D films (Kodak Ultra-

speed) and for digital imaging. For the speed E films
(Kodak Ektaspeed) select three steps shorter exposu-
re time, and for the speed F films (Kodak Insight) fi-
ve steps shorter exposure time.

When using the 30 cm long cone program the values
according to the table given in section 12 “EXPO-
SURE VALUE TABLES” on page 23.

Table 4:Preprogrammed settings values with 20 cm (8”) cone

INCISORS

PREMOLARS 

AND CANINES

MOLARS

OCCLUSAL 

EXPOSURE

kV

time

kV

time

kV

time

kV

time

Adult

Maxilla

60

0.20

63

0.20

63

0.25

70

0.20

Mandible

60

0.16

63

0.16

63

0.20

70

0.20

Child

Maxilla

60

0.12

60

0.16

60

0.20

66

0.16

Mandible

60

0.10

60

0.12

60

0.16

66

0.16

Содержание Prostyle Intra

Страница 1: ...Prostyle Intra USER S MANUAL Publication part number 688092 Version 20 Published 2001 08 En...

Страница 2: ......

Страница 3: ...ers 12 7 POSITIONING THE PATIENT 13 7 1 Positioning the film sensor 13 7 2 Positioning the cone 14 8 TAKING AN EXPOSURE 16 9 EXPOSURE VALUES 17 9 1 Default exposure values 17 9 2 Preprogrammed setting...

Страница 4: ...right to make changes without prior notice COPYRIGHT PLANMECA 2001 08 PUBLICATION PART NUMBER 688092 revision 20 The manufacturer assembler and importer are responsible for the safety reli ability an...

Страница 5: ...dental care professional CAUTION The x ray unit may be dangerous to both patient and operator unless safe exposure values are used and correct operating procedures are observed Prostyle Intra x ray un...

Страница 6: ...s manual 2 SYMBOLS Type B equipment Standard IEC 601 1 Alternating current Standard IEC 417 Attention consult accompanying documents Standard IEC 601 1 Intermediate focal spot Standard IEC 417 Radiati...

Страница 7: ...E USER TO ENSURE THAT ALL LO CAL SAFETY REQUIREMENTS ARE MET Make sure that you are fully acquainted with the app ropriate radiation protection measures and this user s manual before using the unit Al...

Страница 8: ...o the terminal at the under side of the generator box and the other end to the control panel CAUTION Do not connect any other equipment to the control panel s terminal Prostyle Support arm Generator b...

Страница 9: ...tool The stiffness of the tube head horizontal movement has been preadjusted at the factory and can be chan ged by the user if necessary Turn the adjustment screw 0 5 1 rounds clockwise if you want t...

Страница 10: ...an exposure a waiting time starts to flash on the time display which indicates the delay be fore the next exposure can be taken In the digital imaging mode the exposure time is shown with prefix c and...

Страница 11: ...y a third time will recall the default ex posure values The preprogrammed settings can be changed by the user see section 10 PROGRAMMING THE EXPO SURE VALUES on page19 Adult child mode selection key a...

Страница 12: ...icator light will come on when the unit is ready to take an exposure The waiting time between exposures is 15 times exposure time but always at least 6 seconds Exposure key When you take an exposure y...

Страница 13: ...posure and density values on page 19 NOTE There are two sets of default exposure values one for the adult mode and the other for the child mode The unit is always in the adult mode when it is switched...

Страница 14: ...l limator NOTE The exposure values must be selected according to the cone used in the exposure refer to the section 9 EXPOSURE VALUES on p a g e17 NOTE The film holder can be used with the long cone o...

Страница 15: ...he short cone and on the rectangular cone are in line 1 and rotate the cone 180 until the black dot on the rectangular cone and the red dot on the short cone are in line 2 The cone can be now ro tated...

Страница 16: ...will come on The preprogrammed time and kV values appear on the respective displays NOTE The bite wing exposures are taken with the default exposure values Pressing the preprogrammed set tings key a...

Страница 17: ...ne is used for the parallel ing technique Bisecting angle technique The patient holds the film or sensor in place with his finger The x ray beam is directed perpendicularly to wards an imaginary line...

Страница 18: ...n patients upper and lower teeth 7 2 Positioning the cone The angle of the cone is indicated on the scale located on the vertical joint of the tube head Select the cone angle from the table below Tabl...

Страница 19: ...ITIONING THE PATIENT User s manual Position the cone according to the figures below Maxillary anterior Maxillary molar Mandibular anterior Mandibular canine Mandibular molar Maxillary occlusal Mandibu...

Страница 20: ...patient may remain in the radiation area while the exposure is taken Maintain visual con tact with the patient and the unit during the exposure Check that the ready light is on Press and hold the exp...

Страница 21: ...alues on page19 the default exposure valu es are changed automatically Changing the density value will change the selected time value as follows one density step equals to one time step The negative d...

Страница 22: ...UE TABLES on page 23 Changing the density value will change the selected ti me value as follows one density step equals to one ti me step The negative density value shortens the selec ted time value w...

Страница 23: ...and digi tal imaging mode when you exit the programming mode Press and hold down the select key until you have heard two signal tones about 4 seconds to enter the program mode The time display will st...

Страница 24: ...exits the programming mode and the current values will be stored in the memory 10 2 Programming the preprogrammed settings NOTE Two sets of exposure values time kV can be program med for each exposure...

Страница 25: ...select key briefly the kV display will start to flash and the exposure time value is stored in the me mory The kV value can now be changed with the parameter adjustment keys You can now select a new...

Страница 26: ...on Stronger agents can be used for disinfecting the surfa ces We recommend D rr System hygiene FD 322 or respective disinfecting solution Film holder The film holder can be autoclaved up to 145 C or c...

Страница 27: ...8 cone TIME 0 01s 0 02s 0 03s 0 04s 0 05s 0 06s 0 08s 0 10s 0 12s 0 16s 0 20s 0 25s 0 32s 0 40s 0 50s 0 64s 0 80s 1 00s 1 25s 1 60s 2 00s 2 50s 3 20s 70 kV child MAXILLA MANDIBLE I P M O I P M O 66 kV...

Страница 28: ...ANDIBLE I P M O I P M O 63 kV child MAXILLA MANDIBLE I P M O I P M O 60 kV child MAXILLA MANDIBLE I P M O I P M O 57 kV child MAXILLA MANDIBLE I P M O I P M O 55 kV child MAXILLA MANDIBLE I P M O I P...

Страница 29: ...ay tube Anode current mA missing or not in specified limits E 3 3 X ray tube Filament voltage V missing or outside the range too low or too high E 3 4 X ray tube Anode voltage kV missing or below the...

Страница 30: ...ntaining hazardous materials must be disposed of in accordance with waste legislation and instructions issued by the environmental authorities The risks involved and the necessary precautions must be...

Страница 31: ...kV 2 kV Anode current 8 mA Target angle 16 Exposure times 0 01 3 20 sec 23 steps Reference current time product 8 mAs at 70 kV 8 mA 1 sec Lowest current time product 0 08 mAs at 8 mA 0 01 sec Max nom...

Страница 32: ...S 28 Prostyle Intra X ray Unit User s manual Dimensions in mm 1056 220 735 575 575 1150 360 450 320 42 29 9 23 23 46 18 1470 1670 1970 57 5 65 5 77 159 6 25 305 108 4 25 76 3 535 std 21 1 335 opt 13 2...

Страница 33: ...eca Middle East Al Moosa Tower II Sheikh Zayed Road P O Box 28826 Dubai UAE Tel 971 4 3327 668 Head Office Plandent Oyj Asentajankatu 6 00810 Helsinki Finland Tel 358 9 759 05200 Davis Healthcare Serv...

Отзывы: