TECHNICAL SPECIFICATIONS
22 Planmeca ProSensor
User’s Manual
When using the 30 cm long cone program see Planmeca
ProX User’s manual section 17 “EXPOSURE VALUE
TABLES” or select three steps darker density (longer
exposure time).
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV
mA
time
kV
mA
time
kV
mA
time
Adult
Maxilla
60
7
0.200
63
6
0.200
63
6
0.250
Mandible 60
7
0.160
63
6
0.160
63
6
0.200
Child
Maxilla
60
7
0.125
60
7
0.160
60
7
0.200
Mandible 60
7
0.100
60
7
0.125
60
7
0.160
OCCLUSAL
EXPOSURE
ENDODONTIC
BITE-WING
kV
mA
time
kV
mA
time
kV
mA
time
Adult
Maxilla
70
6
0.200
60
7
0.200
60
7
0.200
Mandible 70
6
0.200
Child
Maxilla
66
6
0.125
60
7
0.160
60
7
0.160
Mandible 66
6
0.125