Chapter
G - TROUBLESHOOTING
G-16 Planmeca Proline CC panoramic x-ray
Technical Manual
12.4 Very light radiographs
12.5 Stripes on the film
12.6 Exposed area incorrectly limited
Film has certain tolerance in the holder which causes considerable variations.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Very light radiographs
Film cassette upside down in its
holder.
Primary slot positioned so that the
radiation beam goes trough the
secondary slot.
Secondary slot moved.
Wrong film/intensifying film combi-
nation.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Horizontal stripes
From developing machine.
Primary slot tilted so that part of
the film meets secondary slot.
Patient has jewels, large fillings in
teeth, etc.
Vertical stripes
From rolls of developing machine.
Cassette mechanism gets tuck.
Generator working improperly.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Radiograph incorrectly limited in
vertical direction.
Child limit chosen by accident.
Primary slot limits adjusted incor-
rectly.
Содержание Proline CC
Страница 2: ......
Страница 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 158: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Страница 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 206: ......