In this window the user must mark the corresponding tool position from
which the tool in the collet chuck originates and confirm the entry. If no tool
is located in the collet chuck then the user must enter a “0”. If the entry was
correct then the red symbol appears in green. After the entry, the milling unit
performs the tool change.
8.14 Replacing broken milling cutters
After each tool pick-up and before each tool set-down, the length of the tool
in the collet chuck is measured. The length of the measured tool must lie
within the predefined range.
If a deviation should arise when the tool length is measured, the following
error message appears on the screen of the control PC.
The error that occurs results in the tool being marked as defective in its
position in the tool magazine. In place of the tool symbol, a red warning
triangle now appears, which labels the tool as defective. Until the tool is
replaced and the new tool is measured successfully, the defective tool is
blocked by the control software and cannot be used.
NOTE
After replacing a defective tool:
If the attempt to use a defective tool results in a milling process being
interrupted, the process must be restarted after the defective tool is
replaced!
8 Operating milling unit
58 Planmeca PlanMill 50 S