Planmeca Intra X-ray unit 33
Technical Manual
MECHANICAL ADJUSTMENTS
7.3
Adjusting the stiffness of the tube head’s horizontal axle
Remove the plug from the tube head’s axle and adjust the tightness of the two adjusting
screws evenly (arrows on the figure below).
Figure 10
A
dj_stiffness
.eps
WARNING!
Do not touch the painted screws
ø 2.5 Allen screw
or the earth spring attachment screw.
Содержание intra