29.6.2 After each patient
Steps
1. Remove the used aspirating tips.
2. For hygienic and operational reasons, empty one glass of water (100 -
200 ml) with each suction tube by aspirating water and air through the
suction handpiece. Do this even if only the saliva suction tube has been
used.
3. Wipe the suction handpieces with Planmeca approved surface
disinfectant.
4. Wipe the suction holder / Flexy-holder, including the tablet holder, with
Planmeca approved surface disinfectant.
Results
NOTE
Do not use a spray disinfection solution for the suction arms and holders.
NOTE
Clean the tablet according to the instructions given by the manufacturer.
29.6.3 After each working day
Steps
1. Disinfect the suction system by running Suction cleaning, see section
"Suction cleaning" on page 219.
2. When the suction system has been rinsed, disinfect the suction
handpieces in a washer-disinfector at 93°C, then optionally autoclave
them at 134°C.
3. Flexy-holder: Disinfect the supplementary holders, suction tube holders
(including rollers) and instrument holder in a washer-disinfector at 93°C.
4. Flexy-holder: Wipe the holder for the intraoral scanner with Planmeca
approved surface disinfectant.
29 Cleaning and disinfection
248 Planmeca Compact i
User's manual
Содержание Compact i5
Страница 1: ...PlanmecaCompact i5 user s manual EN 30011782...
Страница 10: ...Table of contents Planmeca Compact i User s manual...
Страница 74: ...5 Replace the suction tubes in the reverse order 11 Suction system 64 Planmeca Compact i User s manual...
Страница 320: ...35 3 5 Side delivery with hanging tube instruments 35 Technical information 310 Planmeca Compact i User s manual...
Страница 325: ......