1. Open the cuspidor door.
2. Pull out the suction tube cleaning holder.
The plug can be detached from the cleaning holder prior to cleaning, but
you can just as well clean the cleaning holder with the plug still attached.
3. Disinfect the suction tube cleaning holder in a washer-disinfector at
93°C.
4. When the suction tube cleaning holder is removed you can access the
separation tank. Clean the separation tank using water and a brush. Be
careful not to dislocate the tubes in the tank.
NOTE
When you place the suction tube cleaning holder back in the cuspidor after
cleaning, make sure it is pushed firmly into its position in the cuspidor.
29.5.4 Coarse filters
Empty the coarse filters daily or when they are full. Replace them monthly.
NOTE
The coarse filter should be emptied into a separate amalgam container.
29 Cleaning and disinfection
242 Planmeca Compact i
User's manual