Chapter G - PARTS REPLACEMENT & REPAIR
Planmeca Compact i G-19
SUCTION SYSTEM RELATED
Technical Manual
4.5 Replacing suction closing membrane
a)
Detach the Microvac II separating tank from the unit as described in section 4.2 “Replacing
the whole PM Microvac II separating tank” on page G-17.
b)
Remove the lid assembly by unscrewing the four M3X16 screws.
c)
Lift off the old suction closing membrane.
d)
Install the new suction closing membrane in reverse order.
NOTE
Care must be taken when positioning the suction closing membrane.
NOTE
Tighten the lid assembly screws crosswise.
MV
ac41298.eps
Screw M3x16
Lid assembly
Suction closing
membrane
Note the positions of the grooves!
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......