Chapter G - PARTS REPLACEMENT & REPAIR
Planmeca Compact i G-17
SUCTION SYSTEM RELATED
Technical Manual
4.2 Replacing the whole PM Microvac II separating tank
WARNING
Used separating tank includes saliva and bacteria. Always wear protective
clothing when handling the separating tank.
a)
Turn off the unit. In the case that suction motor is not controlled by the unit, turn off the suction
motor.
b)
Detach the suction line and the suction filter tubes from the separating tank.
c)
Disconnect the microvac cable from the microvac valve and disconnect the overfill switch
cables.
d)
Disconnect the pulse valve tube from the front of the separating tank.
e)
Lift the separating tank from the socket.
f)
Install the new separating tank in reverse order.
Micro
v
ac
Suction
line tube
Suction filter tube
Microvac II socket
Pulse valve tube
Microvac cable
Overfill
switch cables
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......