30 Technical information
30.1 Technical specifications
Original manufacturer
PLANMECA Oy, Asentajankatu 6, 00880 Helsinki, FINLAND
Phone: +358 20 7795 500, fax: +358 20 7795 555, www.planmeca.com
Colours
Painted parts:
RAL-9016
Upholstery colours:
Please consult your dealer for availability
Mechanical dimensions
Installed:
(H x D x W) 1210 mm x 963 mm x 1990 mm
(see template for details)
Weight
150 kg (331 lbs)
Maximum allowed load on patient seat
Maximum allowed patient weight 185 kg (407 lbs) and accessories load of
15 kg (33 lbs)
Environmental conditions
Transportation conditions
Temperatures:
-20°C to +60°C (-4°F to +140°F)
Relative humidity:
5% RH to 95% RH; non-condensing humidity
Air pressure:
700 hPa to 1060 hPa (10 psi to 15 psi)
Storage conditions
Temperatures:
-5°C to +60°C (+23°F to +140°F)
Relative humidity:
5% RH to 95% RH; non-condensing humidity
Air pressure:
700 hPa to 1060 hPa (10 psi to 15 psi)
If the unit has been stored at temperatures below +10 °C (+50 °F) for more
than a few hours, time must be allowed for the unit to reach room
temperature in the original packing before connecting the unit to the mains
voltage.
Operating conditions
Temperatures:
+15°C to +35°C (+59°F to +95°F)
Relative humidity:
5% RH to 95% RH; non-condensing humidity
Air pressure:
800 hPa to 1060 hPa (12 psi to 15 psi)
Altitude:
< 2000 m (less than 1.25 miles)
Mains voltage and frequency
Mains voltage setting:
100V~
115V~
220-240V~
Mains frequency:
50 or 60 Hz
Fuse rating and type
30 Technical information
User's manual
Planmeca Compact i 169
Содержание Compact i Classic v2
Страница 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Страница 8: ...Table of contents Planmeca Compact i User s manual...
Страница 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Страница 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Страница 185: ...30 Technical information User s manual Planmeca Compact i 177...
Страница 189: ......