26.2 Units with VS/A compatible suction system
1. Bowl filter
Empty daily and replace when necessary.
2. Coarse filters
Empty daily and replace weekly.
3. Fine filter
Empty when full. Refer to the instructions supplied with the DÜRR
amalgam separator.
26.3 Oil collector
26.3.1 Instrument console with balanced instrument arms
The oil collector underneath the instrument console has to be emptied and
cleaned monthly. Remove the oil collector by turning it counter-clockwise as
shown in the picture below.
26 Maintenance
User's manual
Planmeca Compact i 139