INSTALLATION:

The Planet Model ULED arm is designed to be mounted above the immediate work area. Standard mounting fi xtures for Planet Model ULED are: 
Ceiling Mount & Wall Mount (specialised mountings are available on application). In all cases the suggested mounting height (measured at swivel 
head) is two metres from fl oor level. (refer  Fig 1)
To ensure correct balance of the Model ULED movement the lamp must be mounted with the stem, around which it rotates, vertical at all 
times.
(refer  Fig 1)

  

WARNING

  The qualifi ed electrician who is responsible for the installation of the ULED must include an isolation switch( all -pole discon-

nection ) in the fi xed wiring of the ULED and a cable tie near the terminal block of the primary(240V) wiring (Phase and Neutral) to secure the 
cables.

Figures 2 - 4 Refers to wall mount options, Figures 6 - 8 Refers to ceiling mount options, Figures 9 - 11 Refers to maintenance procedures 

5.0  WALL MOUNT:

 

  

WARNING

   

Ensure before commencing the installation, that the MAINS power circuit is isolated (off).

CHECK AND ENSURE THE FOLLOWING BEFORE COMMENCING INSTALLATION:
•   

You have the appropriate mounting template.

•   

You have appropriate tools for the job.

•   

Confi rmed mains cable entry is located appropriately. 

•   

The mounting structure is suitable. 

•   

The structural loadings and approx product weights (refer Fig12) 

•   

You have appropriate fi xing attachments. e.g. Dyna bolts, large screws 

5.1.0   Using appropriate wall template (template 1) mark mounting holes and mains cable entry. Make sure the bottom of the template   
 

 

wall mount is 2055mm from the fl oor to maintain 2000mm clearance at the bottom of the swivel. (ref Fig 1)

5.2.0    Drill 4 mounting holes as shown on template 1.

Fig: 2

Fig: 10

Fig: 11

Depress button to 
remove or replace light head 

 Lock Nut

left hand
(heavy) spring tube

Spring tension nut

  

WARNING

  

UNDER NO CIRCUMSTANCES SHOULD THE 

SPRING TENSION NUTS (TWO) BE COMPLETELY UNDONE

Classifi cation: Non invasive medical 
device, as per label/s below.

Содержание ULED

Страница 1: ... must be stored warehoused in a dry place and as per the same environmental conditions mentioned above 2 2 0 Disposal and Recycling Information The ULED light has been designed taking into account what the manufacturing components are made out of This is to ensure when it reaches its end of life the disposal method with the lowest least negative impact on human health and the environment is select...

Страница 2: ...11 Refers to maintenance procedures 5 0 WALL MOUNT WARNING Ensure before commencing the installation that the MAINS power circuit is isolated off CHECK AND ENSURE THE FOLLOWING BEFORE COMMENCING INSTALLATION You have the appropriate mounting template You have appropriate tools for the job Confirmed mains cable entry is located appropriately The mounting structure is suitable The structural loading...

Страница 3: ...ists and descriptions will be made available to assist in this process The following maintenance procedures performed as required should ensure your Planet Model ULED provides many years of effective service 9 1 Cleaning Carefully remove any loose deposits with a damp sponge Use a soft brush non abrasive or cloth and or a mild household detergent solution to remove dust and other deposits Do not u...

Страница 4: ...ove plastic side cover and screw 10 from continuous rotation device Slide push cable from the light head through tube 11 and out the side cover of the continuous rotation device 12 Ensure the grommet 13 is located correctly Trim and strip wires to appropriate length Attach swivel with bolt and dome nut and tighten 14 When attaching the swivel to the bottom spigot ensure the spigot does not slide u...

Отзывы: