background image

Model ‘ULED’ Spring Balanced Examination Light

 - 

Installation & Maintenance  

Instructions

1.0  

INTRODUCTION

This installation and instruction manual describes how to install, maintain and operate the Planet Model 
‘ULED’ examination light. Prior to installation carefully read through all the instructions enclosed in this 
manual.
If there are any concerns or you do not understand, please contact customer service at Planet Lighting.
 

 

  WARNINGS

Only qualifi ed electricians can install the Planet model ‘ULED’ light. It is to be direct wired only. Installation 
performed by unauthorised personnel could result in personal injury and or equipment damage. Under 
no circumstances should any modifi cations or adjustments, other than those detailed under section  9.0 
Maintenance, be made to the Planet Model ULED.
      
2.0  

Product Information

  The Planet ULED is a state of the art examination light which utilises cutting edge Solid State Lighting technology 
combined with the proven Planet Model U spring arm and bracket system. Designed and manufactured in Australia by 
Planet Lighting (established 1911) an R&D driven, exporting designer and manufacturer with over 6 years experience 
in the design, manufacture and supply of Solid State Lighting devices worldwide.
   The Planet ULED provides an improved working environment, signifi cantly reducing on-going maintenance, 
with low operational costs and offers a “greener” alternative to similar products.
The Planet model ULED allows the accurate positioning of the lamp head above any work surface. It is designed to be 
mounted on a wall or ceiling and be pulled down from above when required, remaining there until re - positioned.
At Planet we take great pride in the design and production of our lights and we are sure you will derive satisfaction 
from owning and using our products. When installed and maintained in accordance with the instructions contained in 
this service leafl et, this product will provide many years of trouble free operation.
Should you have any enquiry, comment or special lighting need we would like to hear from you. We back our products 
with pride, service and a fi ve year warranty.

 

2.1.0  

Environmental Considerations  

 2.1.1 Packaging
 

   

I box containing a complete ULED lamp with Installation and users manual.

 2.1.2 Transport 
 

   

transport to be carried out by road, while maintaining the following environmental conditions.

 

   

Temperature: -15/+60 degrees Celsius 

 

   

Humidity : 10/75%

 

   

Atmospheric pressure(h/Pa) 500/1060

 2.1.3 Storage
 

   

The devices packaged must be stored (warehoused) in a dry place and as per the same  

 

   

environmental conditions mentioned above.

 

2.2.0 

Disposal and Recycling Information

 

   

The ULED light has been designed, taking into account what the manufacturing components are

  

   

made out of. This is to ensure when it reaches its end of  life, the disposal method  

 

 

   

with the lowest/ least negative impact, on human health and the environment is selected.  

 

   

Unlike the sources of traditional light, LED’s do not contain mercury, toxic gases, fi laments or  

 

   

fragile parts. Due to the fi xture being over 95% recyclable and there being no Bulbs/ 

 

 

   

Globe changing during the ULED life of 50,000+ hours your environmental footprint will be  

 

 

 signifi 

cantly 

reduced. 

  

 

   

When your ULED light reaches its end of life, contact Planet lighting or our local agents to learn 

 

   

about recycling options.The ULED light must be disposed of properly according to local laws and,  

  

  regulations. 

 

3.0  

PLANNING

Prior to installing the Planet Model ‘ULED’ Ceiling and Wall mounts the following items should be checked and con-
fi rmed.

 

3.1.0     

 Architectural considerations

 

3.1.1      

Ensure appropriate mounting structure behind the wall/ceiling

 

3.1.2     

For structural loadings and appox product weights (refer Fig12)

 

3.1.3     

To confi rm ceiling/wall height measurements at the appropriate nominal recommended 

 

   

height from the fl oor refer Fig 1 (back page) 

 

3.2.0    

Electrical considerations

 

3.2.1   

The Planet Model  ‘ULED’ wall and ceiling mount are designed to be direct wired only

 

3.2.2   

240V AC supply is available at the connection points

Made in Australia Exported to the World

Tamarind Drive Bellingen NSW 2454 Australia
Phone + 61 (0)2 6659 5600     

Fax + 61 (0)2 66552121

[email protected]           

www.planetlighting.com

Customer Service can be contacted on the phone number below Mon - Friday   9 am - 5 pm

Fig: 12

Recommended

height: 2000 mm

to bottom of lamp pivot

 Standard ceiling fix

Suspended ceiling fix

Structural ceiling

Recommended

height: 2055 mm

to bottom of

mounting plate

 Wall fix

Fig: 1

Recommended

height: 2000 mm

to bottom of  swivel head

Ensure

level

Ensure

level

Recommended

height: 2055 mm

to bottom of

mounting plate

610 NOM (= ~16 Nm)

445 NOM (= ~12 Nm)

RELATIVE VERTICAL

MASS: 2.7kg

POSITION SHOWN HERE
IS ESTIMA TE ON LY.

(ARM+HEAD ON LY)

Note:
For the purposes of quoting minimum structural requirements of mountings, torque-load of 12 Newton Metres 
plus lever-arm effect of length of ceiling-mount tube or wall mounting must be accommodated (plus factors 
of safety and dynamic loads). Vertical dead load (IE No dynamic or external loads such as operators pulling or 
pushing fi tting are included in this value.)

Static weights of complete fi xture (Mount, Arm & Light Head)

 

Wall Mount  

-  

5.5kg

 Ceiling 

Mount  -  6.0kg

Содержание ULED

Страница 1: ... must be stored warehoused in a dry place and as per the same environmental conditions mentioned above 2 2 0 Disposal and Recycling Information The ULED light has been designed taking into account what the manufacturing components are made out of This is to ensure when it reaches its end of life the disposal method with the lowest least negative impact on human health and the environment is select...

Страница 2: ...11 Refers to maintenance procedures 5 0 WALL MOUNT WARNING Ensure before commencing the installation that the MAINS power circuit is isolated off CHECK AND ENSURE THE FOLLOWING BEFORE COMMENCING INSTALLATION You have the appropriate mounting template You have appropriate tools for the job Confirmed mains cable entry is located appropriately The mounting structure is suitable The structural loading...

Страница 3: ...ists and descriptions will be made available to assist in this process The following maintenance procedures performed as required should ensure your Planet Model ULED provides many years of effective service 9 1 Cleaning Carefully remove any loose deposits with a damp sponge Use a soft brush non abrasive or cloth and or a mild household detergent solution to remove dust and other deposits Do not u...

Страница 4: ...ove plastic side cover and screw 10 from continuous rotation device Slide push cable from the light head through tube 11 and out the side cover of the continuous rotation device 12 Ensure the grommet 13 is located correctly Trim and strip wires to appropriate length Attach swivel with bolt and dome nut and tighten 14 When attaching the swivel to the bottom spigot ensure the spigot does not slide u...

Отзывы: