Model ‘ULED’ Spring Balanced Examination Light
-
Installation & Maintenance
Instructions
1.0
INTRODUCTION
This installation and instruction manual describes how to install, maintain and operate the Planet Model
‘ULED’ examination light. Prior to installation carefully read through all the instructions enclosed in this
manual.
If there are any concerns or you do not understand, please contact customer service at Planet Lighting.
WARNINGS
Only qualifi ed electricians can install the Planet model ‘ULED’ light. It is to be direct wired only. Installation
performed by unauthorised personnel could result in personal injury and or equipment damage. Under
no circumstances should any modifi cations or adjustments, other than those detailed under section 9.0
Maintenance, be made to the Planet Model ULED.
2.0
Product Information
The Planet ULED is a state of the art examination light which utilises cutting edge Solid State Lighting technology
combined with the proven Planet Model U spring arm and bracket system. Designed and manufactured in Australia by
Planet Lighting (established 1911) an R&D driven, exporting designer and manufacturer with over 6 years experience
in the design, manufacture and supply of Solid State Lighting devices worldwide.
The Planet ULED provides an improved working environment, signifi cantly reducing on-going maintenance,
with low operational costs and offers a “greener” alternative to similar products.
The Planet model ULED allows the accurate positioning of the lamp head above any work surface. It is designed to be
mounted on a wall or ceiling and be pulled down from above when required, remaining there until re - positioned.
At Planet we take great pride in the design and production of our lights and we are sure you will derive satisfaction
from owning and using our products. When installed and maintained in accordance with the instructions contained in
this service leafl et, this product will provide many years of trouble free operation.
Should you have any enquiry, comment or special lighting need we would like to hear from you. We back our products
with pride, service and a fi ve year warranty.
2.1.0
Environmental Considerations
2.1.1 Packaging
I box containing a complete ULED lamp with Installation and users manual.
2.1.2 Transport
transport to be carried out by road, while maintaining the following environmental conditions.
Temperature: -15/+60 degrees Celsius
Humidity : 10/75%
Atmospheric pressure(h/Pa) 500/1060
2.1.3 Storage
The devices packaged must be stored (warehoused) in a dry place and as per the same
environmental conditions mentioned above.
2.2.0
Disposal and Recycling Information
The ULED light has been designed, taking into account what the manufacturing components are
made out of. This is to ensure when it reaches its end of life, the disposal method
with the lowest/ least negative impact, on human health and the environment is selected.
Unlike the sources of traditional light, LED’s do not contain mercury, toxic gases, fi laments or
fragile parts. Due to the fi xture being over 95% recyclable and there being no Bulbs/
Globe changing during the ULED life of 50,000+ hours your environmental footprint will be
signifi
cantly
reduced.
When your ULED light reaches its end of life, contact Planet lighting or our local agents to learn
about recycling options.The ULED light must be disposed of properly according to local laws and,
regulations.
3.0
PLANNING
Prior to installing the Planet Model ‘ULED’ Ceiling and Wall mounts the following items should be checked and con-
fi rmed.
3.1.0
Architectural considerations
3.1.1
Ensure appropriate mounting structure behind the wall/ceiling
3.1.2
For structural loadings and appox product weights (refer Fig12)
3.1.3
To confi rm ceiling/wall height measurements at the appropriate nominal recommended
height from the fl oor refer Fig 1 (back page)
3.2.0
Electrical considerations
3.2.1
The Planet Model ‘ULED’ wall and ceiling mount are designed to be direct wired only
3.2.2
240V AC supply is available at the connection points
Made in Australia Exported to the World
Tamarind Drive Bellingen NSW 2454 Australia
Phone + 61 (0)2 6659 5600
Fax + 61 (0)2 66552121
www.planetlighting.com
Customer Service can be contacted on the phone number below Mon - Friday 9 am - 5 pm
Fig: 12
Recommended
height: 2000 mm
to bottom of lamp pivot
Standard ceiling fix
Suspended ceiling fix
Structural ceiling
Recommended
height: 2055 mm
to bottom of
mounting plate
Wall fix
Fig: 1
Recommended
height: 2000 mm
to bottom of swivel head
Ensure
level
Ensure
level
Recommended
height: 2055 mm
to bottom of
mounting plate
610 NOM (= ~16 Nm)
445 NOM (= ~12 Nm)
RELATIVE VERTICAL
MASS: 2.7kg
POSITION SHOWN HERE
IS ESTIMA TE ON LY.
(ARM+HEAD ON LY)
Note:
For the purposes of quoting minimum structural requirements of mountings, torque-load of 12 Newton Metres
plus lever-arm effect of length of ceiling-mount tube or wall mounting must be accommodated (plus factors
of safety and dynamic loads). Vertical dead load (IE No dynamic or external loads such as operators pulling or
pushing fi tting are included in this value.)
Static weights of complete fi xture (Mount, Arm & Light Head)
Wall Mount
-
5.5kg
Ceiling
Mount - 6.0kg