SpectraVideo Camera User’s Manual
Appendices
43
13.3.2.
DSP
C
ONTROL
O
UTPUTS
SpectraVideo Camera Control commands
Clamp Sample from DSP
Shutter
Shutter open during Exposure period closed during readout. This signal is used
as Shutter Sync Out
Frame
Blanking
High for flush period and exposure period, low for readout.
Line
Blanking
High for row shift period of each line. The PCI I/O board blanks the first LB
pulse since data is not meaningful until after the first line is written to line
buffer.
An additional line must be read from CCD so that that line will be read from PCI
I/O board line buffer to PC ram.
Data Valid
Rising edge pulse that latches data word to pixel or line buffer in I/O board.
Data
Output
Enable
In SpectraVideo cameras with multiple analog channels, a group of 4 pulses
enable individual Pixel Data latches for transfer to the I/O board. This pulse is
internal to the camera.
Содержание SPECTRAVIDEO