Specifications
Electrical Requirements
100~240VAC, 50/60Hz ; 12VDC
Power Consumption
< 15 watt
Dimensions
3.4”(D)×2.2”(W)×4.9”(H)
3.4”(D)×2.2”(W)×6.9”(H)(incl. battery )
Weight
370 g (0.81 lbs)
615 g (1.35 lbs)(incl. battery )
Sound Pressure Level
53dB(A)
Max. Comp. Pressure
30 Psig or greater
Neb. Operating Pressure
10~14 Psig
Operating Flow
2 LPM
Particle Size Less Than (MMAD)
5 µm
Ampoule Capacity
6 ml
Intermittent Use
20 min ON/40 min OFF
Operating Temperature
10
~ 40
(50
to 104
)
℃
℃
℉
℉
Operating Humidity
Up to 95% non-condensing
Storage/Transport Temperature Range
-20
~ 60
(
℃
℃
-4
to 140
)
℉
℉
Storage/Transport Humidity
Up to 95% non-condensing
(1). Conditions may vary based on altitude above sea level, barometric pressure and temperature.
(2). This product complies with the following standards: C22.2 No 601.1-M90,
UL Std. No 2601-1 and IEC/EN 60601-1.
(3). For electromagnetic compatibility the standards are based as follow: CSA 601.1-2
and IEC/EN 60601-1-2.
(4). Protection against electric shock is classified as
Type II
.
(5). Degree of protection against electric shock is classified as
Type BF
.
(6). Mode of operation belongs to intermittent mode.
(7). Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
(8). Degree of protection against ingress of water is rated as IPX0 (Device is not water proof).
5