Pintech PT-710 Скачать руководство пользователя страница 4

These characteristics apply to an adjusted PT-710 AC/DC Current Probe installed on an
oscilloscope of any brand. The oscilloscope must be warmed up for at least 20 minutes and be in
an environment with the
temperature at 10

~30

and the humidity at 0~80.

Environmental Characteristics

Basic Operation:

Before using the probe, the batteries must be installed.

WARNING!

Do not clamp the probe onto circuits with voltages greater than 600 VAC. Personal injury or
damage to the probe may result. Always connect the PT-710 current probe output to the
instrument before clamping onto the circuit under test.

1. First connect the current probe BNC connector to BP-250 (double BNC connection cable) then
connect to oscilloscope input. Start by setting the oscilloscope voltage input channel to DC volts,
and the voltage sensitivity scale to 5m V/div.
2. Move the OFF/ Range switch to the 100mV/A or 10mV/A position to turn on the probe. (

The

PT-710 current probe has a green LED power/battery indicator. If the LED does not light, replace
the battery or use specified power adaptor.)
3. Use the ZERO adjustment to zero or offset the probe output detection of residual magnetic DC
charges.
4. Connect the probe to the circuit by opening the jaws and clamping around the conductor.

NOTE. Clamping around both the “hot” and neutral wires may give you a zero

reading. (Remember to unclamp the probe from the conductor before disconnecting

it from your meter or instrument).

Size

231 mm x67 mm x 36 mm

Maximum Conductor
Size

11 mm

Cable Length

200 cm

Weight

310 g (without battery)

Temperature

Working

Storage

0°C to +50°C
(+32°F to +122°F)

-20°C to +80°C
(-4°F to +176°F)

Humidity

0°C to 40°C, 95% humidity
40°C to 50°C, 45% humidity

Pollution Degree

2

Содержание PT-710

Страница 1: ...m service procedures To Avoid Fire or Personal Injury Connect and Disconnect Properly Connect the probe output to the measurement instrument before connecting the probe to the circuit under test Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement Instrument Observe All Terminal Ratings To avoid fire or shock hazard observe...

Страница 2: ...tatements identify conditions or practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could result in damage to this product or other property Terms on the Product These terms may appear on the product DANGER Indicates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately ...

Страница 3: ...t needed for AC measurements unless your instrument cannot isolate a DC component if present OFF Range switch Slide the switch from OFF to either the 100 mV A or 10mV A range When either range is selected the probe is turned on and the green battery indicator lights If the indicator does not light see Battery Notes and Battery Installation Battery indicator The green battery indicator lights when ...

Страница 4: ...pe voltage input channel to DC volts and the voltage sensitivity scale to 5m V div 2 Move the OFF Range switch to the 100mV A or 10mV A position to turn on the probe The PT 710 current probe has a green LED power battery indicator If the LED does not light replace the battery or use specified power adaptor 3 Use the ZERO adjustment to zero or offset the probe output detection of residual magnetic ...

Страница 5: ...quency current the momentary peaks may be quite high 7 PT 710 a multifunctional current probe When connecting to a digital meter use the recommended MT 246N BNC to banana adapter Connect the black lead to the meter COM black letters on the meter and the red lead to the VΩ input red letters on the meter 8 To measure only AC current set the meter to measure AC volts 9 To measure DC current set the m...

Страница 6: ...age V Rating Range 10mV A Range 100mV A Max Working voltage V Max floating voltage V DC 100 10 600 600 DC peak AC 100 10 600 600 AC peak 100 10 600 600 AC peak peak 200 20 1200 RMS CAT III 70 7 600 600 RMS CAT II 70 7 600 600 RMS CAT I 70 7 600 600 Voltage and current ratings 8 ...

Отзывы: