PN 350-007660-07 

 

P a g e

 | 

4

 

 

 

 

3.

 

Holding the digital force adapter, rotate the holder screw until it touches the top of the clamp arm. 
At this point, there should be about a 1/16

 

inch gap between the top of the digital force adapter and 

the clamp arm.  

IMPORTANT:  The connection of the wire to the load cell is fragile. 

Gently guide the movement of 

the wire, being careful not to pull it. 

       

 

 

4.

 

Using the provided plastic hooks, secure the cord from the load cell assembly along the horizontal 
arm of the clamp. Place the load cell cord into the eye of the hook and snap the hook closed. 
Remove the adhesive backing and place on the horizontal arm portion of the clamp. 
 

 

 
5.

 

The sample tip is removable. Screw onto the load cell assembly an appropriate press tip (swivel, flat, 
or concave) for the sample type to be analyzed. The various press tips are detailed in the MIRacle 
manual and in the Digital Force Adapter Usage section on page 8 of this manual.  

1/16 inch gap 

Содержание MIRacle

Страница 1: ...Digital Force Adapter for High Pressure Clamp Installation and User Guide...

Страница 2: ...cument contains confidential or proprietary information of PIKE Technologies Inc Neither this document nor the information herein is to be reproduced distributed used or disclosed either in whole or i...

Страница 3: ...lation for MIRacle 3 Installation for GladiATR or GladiATR Vision 5 Digital Force Adapter Usage 8 Choosing a Press Tip 8 Clamp Operation 9 Zero Adjustment of Digital Force Adapter 9 Application 10 Spe...

Страница 4: ......

Страница 5: ...he High Pressure Clamp of either the MIRacle or GladiATR It consists of an embedded load cell which exhibits high linearity and exceptional reproducibility To accommodate various sample types the adap...

Страница 6: ...e contact PIKE Technologies Inc immediately if any parts have been omitted Packing List Manual PN 350 007660 MIRacle GladiATR Load Cell Assembly with Tip LCD Monitor Quantity 1 Quantity 1 Quantity 1 P...

Страница 7: ...clarity Installation for MIRacle 1 Remove the press tip rod from the MIRacle High Pressure Clamp by unscrewing the rod from the top of the horizontal clamp arm 2 Hold the digital force adapter assembl...

Страница 8: ...ll it 4 Using the provided plastic hooks secure the cord from the load cell assembly along the horizontal arm of the clamp Place the load cell cord into the eye of the hook and snap the hook closed Re...

Страница 9: ...is not connected to the LCD monitor prior to applying power erroneous readings are displayed 7 Choose an appropriate electrical outlet adapter and attach it to the power cord Plug the power cord into...

Страница 10: ...tip until finger tight Exhibit care not to excessively twist the cord attached to the load cell assembly To minimize cord twisting allow the cord to hang vertically during installation 4 To reinstall...

Страница 11: ...rtion of the clamp 7 Insert the connector of the load cell assembly and the power cord into the back of the LCD monitor The appropriate location of each connector is clearly labeled If the load cell i...

Страница 12: ...to the crystal surface This tip allows for better positioning and optimal contact of thin materials with the crystal surface The swivel tip is used with irregularly shaped samples films and semi rigid...

Страница 13: ...he control knob turning clockwise until the desired pressure shown on the digital LCD monitor is achieved 4 Collect the sample spectrum Zero Adjustment of Digital Force Adapter The Digital Force Adapt...

Страница 14: ...e Adapter Spectra were collected at two force settings 10 and 20 kgf The results show band intensity increases with increasing pressure Using the Digital Force Adapter provides control over applied fo...

Страница 15: ...just screw located on the right side of the LCD monitor Service and Repairs For service and repairs please contact PIKE Technologies Inc to obtain a Returned Merchandise Authorization RMA number When...

Страница 16: ...wheeled bin is a clear reminder that the product must NOT be disposed with household waste It is the responsibility of the buyer to discard the product in accordance with Federal regional and local en...