Precautions
Cautions for Measurement of Samples in a Container
This accessory scans samples in glass vials or quartz cuvettes. The spectrum of the vial or cuvette
superimposes the sample spectrum. When you use other glass containers for analysis, you have to
investigate influences that the containers may give to the spectra and result before starting the
analysis.
SAFETY Precaution
The maximum temperature of the accessory can reach 130 °C. Please allow enough time for the
metal parts to cool down before working with the Falcon components or removing it from the
spectrometer. Remove heated sample vials or cuvettes by using tweezers. Do not to touch
heated vial or cuvette sample holders.