Tergeo plasma cleaner operator’s manual
Revision: B3, Year 2020
5)
Vacuum pump power outlet port on the plasma machine: For North America and Japan, type:
NEMA 5-15; Rating: 15A, 125VAC. For regions that use 220V to 230VAC, Type: IEC 320 C13,
Rating: 10A, 250VAC.
6)
The maximum current for the vacuum pump connected to the pump power outlet should be
less than 7 amps.
7)
Compatible power cord for both the plasma system and the vacuum pump: For countries and
regions use 100~110VAC, voltage rating on the power cord must be higher than 110VAC, current
rating should be equal or higher than 10 amps; For countries and regions uses 220~230VAC,
voltage ratings on the power cord should be higher than 230VAC, current rating should be equal
or higher than 10amps.
5.4 Environmental conditions
1.
The recommended condition for operation: temperature range is 0-30C, humidity: 10% -
70% RH.
2.
The recommended conditions for transportation and storage: temperature range is -20-
60C, humidity: 10% - 90% RH.
RF high voltage exists within controller and plasma source
enclosure. Please do not open the enclosure whenever RF
output is ON.
Read this manual carefully before operating the plasma
cleaner for the first time. Failure to follow the safety
precautions may result in serious damage to the system or
may cause personal injury. PIE Scientific LLC is not liable
and will not honor the warranty if the applicable safety
precautions are ignored by the users.
The equipment and the vacuum pump could weight more than
40 lbs. Two persons are required to lift the equipment to avoid
back injury or dropping the equipment to the ground and
causing personal injury.