To start the Automatic Stepping

Start automatic stepping  or ramping by placing the EZ-CHECK 
Switch into the HI or LO position then press and hold the 

 

DIAL KNOB for 6 seconds  (the word STORE will appear on the 
display after 3 seconds and continue to press the EZ-DIAL 
KNOB) until the word STEPPING appears on the display.  The 
word STEPPING will appear on the display anytime the selected 
automatic function is running. Stop the stepping by again 
pressing and holding the 

 DIAL KNOB for 3 seconds. 

BACKLIGHT

 - If BACKLIGHT is ON the backlight will light all the 

time the unit is powered up. For maximum battery life turn the 
backlight off when using the calibrator in areas with enough 
ambient light to read the display.

To change the Automatic Stepping settings

Double click the 

 DIAL KNOB at any time the unit is on and 

the following typical display (will be different for each FUNCTION) 
will appear for 15 seconds:

Turn the 

 DIAL KNOB to move to the second menu page.

Turn the 

 DIAL KNOB to move through the menu. Press the 

 

DIAL KNOB to toggle between OFF and ON or to change the 
STEPS and the STEP TIME settings. These settings are 
remembered even with the power off. 

EXIT MENU

 - exits this menu immediately and saves any 

changes. Menu will automatically exit after 15 seconds of 
inactivity.

AUTO OFF

 - If AUTO OFF is ON, the unit will turn off after 30 

minutes of inactivity to save battery life. If AUTO OFF is OFF the 
unit will stay on until the POWER SWITCH is moved to the off 
position.

>  EXIT   

 

 

      

 

STEPS 

 

 

        3

 

STEP TIME   

        5

 

AUTO OFF   

     ON

 

BACKLIGHT  

     ON

STEPS

 - pressing the knob will cycle through 2, 3, 5, 11 and 

RAMP. The endpoints of the steps or ramp are based on the 
values stored in the 

HI

 and 

LO

 EZ-CHECK outputs. 

2 steps

 will automatically switch between the values stored in 

the HI & LO EZ-CHECK (0 & 100%). 

3 steps

 between the HI, Midpoint and LO EZ-CHECK (0, 50 

& 100%). 

5 steps

 between the HI and LO EZ-CHECK in 25% 

increments (0, 25, 50, 75 & 100%).

11 steps

 between the HI and LO EZ-CHECK in 10% 

increments (0, 10, 20...80, 90 &100%).

RAMP

 continuously ramps up and down between the HI and 

LO EZ-CHECK outputs.

STEP TIME

 - pressing the knob will cycle through 5, 6, 7, 8, 9, 

10, 15, 20, 25, 30 and 60 seconds. 

>  EXIT   

 

 

      

 

FUNCTION   

      mA

 MODE   

 

 

SOURCE

 

UNITS 

 

 

      mA

 HART 

250

Ω

  

      ON

Operating Instructions

Double Click Menu - STEPPING, AUTO OFF & BACKLIGHT

Operating Instructions

Double Click Menu - STEPPING, AUTO OFF & BACKLIGHT

Содержание PIECAL 820

Страница 1: ...Practical Instrument Electronics Please contact an authorized sales representative at 800 346 4620 or order at www Instrumentation com ...

Страница 2: ...imulate 2 Wire Transmitters Power up transmitters loops Calibrate in temperature with 0 1 C F resolution Types J K T E R S B N G C D L J DIN U T DIN P II Pt 100 Ohm 3850 3902 3916 3926 1000 Ohm 3850 RTD Copper 10 50 Ohm Nickel 100 and 120 Ohm RTD Checkout flow and vibration systems Source read frequency from 0 to 2000 CPM Counts Per Minute 0 00 to 999 99 Hz 0 0 to 9999 9 Hz and 0 000 to 20 000 kHz...

Страница 3: ...ged externally are available STORING HI and LO EZ CHECK Source Outputs Speed up your calibration by storing Span Zero output setting for instant recall with the EZ CHECK switch 1 Store your high SPAN output temperature by moving the EZ CHECK switch to the HI position and turning the EZ Dial knob until the desired output value is on the display Press and hold the EZ Dial knob until STORED appears t...

Страница 4: ...igit steps at a time Push down and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for two more seconds to start the automatic ramping READ Press and hold to transfer the current temperature into the EZ Check MIN MAX points This clears the MIN MAX readings which will update as...

Страница 5: ...T FUNCTION T C UNITS C F T C TYPE J K T E R S B G C D L U P COLD JUNC ON EXIT FUNCTION RTD UNITS C RTD Pt 100 3850 RTD Types WIRE MODE 234W EXIT FUNCTION RTD UNITS C RTD Pt 100 3850 RTD Types WIRE MODE 3W 2W 4W RTD Types Pt 100 a 3850 a 3902 a 3916 a 3926 Pt 1000 a 3850 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Ni 110 a 5801 Source mA Simulate 2 Wire Transmitters Read mA Power Measure Transmitters S...

Страница 6: ... change the STEPS and the STEP TIME settings These settings are remembered even with the power off EXIT MENU exits this menu immediately and saves any changes Menu will automatically exit after 15 seconds of inactivity AUTO OFF If AUTO OFF is ON the unit will turn off after 30 minutes of inactivity to save battery life If AUTO OFF is OFF the unit will stay on until the POWER SWITCH is moved to the...

Страница 7: ...ion on the EZ CHECK switch The output is adjusted in 0 01 mA 0 1 increments by turning the knob Press and turn the knob for faster dialing with 1 00 mA 10 0 increments To Sensor Typical 2 Wire Transmitter Disconnected IN REF OUT Receiver Powers External 2 Wire Transmitter Power Supply 2 to 100 VDC SOURCE HI SET LO MAX READ MIN READ 2 Wire SIM mA 2 Wire SIM Percent of 4 to 20 mA Choose this functio...

Страница 8: ... hold the knob to clear the MAX and MIN readings All milliamp connections Typical 2 Wire Transmitter IN REF OUT Transmitter Input Sensor Process Signal Simulated Input SOURCE HI SET LO MAX READ MIN READ Power Measure mA Power Measure Percent of 4 to 20 mA Choose this function to simultaneously supply power to a 2 Wire Transmitter while displaying the 4 to 20 mA output of the transmitter Move the p...

Страница 9: ...ECK switch The output is adjusted in 0 01 mV or V increments by turning the knob Press and turn the knob for faster dialing with 1 00 mV or V increments Process Calibrator mA V TC Ω RTD Hz SOURCE HI SET LO OFF MAX READ MIN READ OVERLOAD Hz SYNC Double Click Menu Push Turn for Fast Dialing Push Hold to Store Step READ mV 8 20 Voltage Output Signal Controller Transmitter Power Supply Read mV V Choos...

Страница 10: ...uble Click Menu Turn the knob to scroll through the settings and press the knob to make your selection Select T C for the FUNCTION F or C for the UNITS T C Type J K T E R S B N G C D L J DIN U T DIN or P Platinel II and COLD JUNC ON or OFF ON is the default Connect the PIECAL 820 to the inputs of the device being calibrated using the proper type of thermocouple wire via the miniature thermocouple ...

Страница 11: ...he knob to clear the MAX and MIN readings Resistance Source Choose this function to provide a simulated resistance into any device that measures resistance Move the power switch to SOURCE then Double Click the EZ DIAL knob to get into the Double Click Menu Turn the knob to scroll through the settings and press the knob to make your selection Select OHMS for the FUNCTION 400Ω or 4000Ω for the RANGE...

Страница 12: ...nput devices that measure RTD sensors Move the power switch to SOURCE then Double Click the EZ DIAL knob to get into the Double Click Menu Turn the knob to scroll through the settings and press the knob to make your selection Select RTD for the FUNCTION F or C for the UNITS and RTD Choose from one of Platinum 100Ω or 1000Ω Copper 10Ω or 50Ω Nickel 120Ω or 110Ω curves Note Pt 100Ω 3850 is the most ...

Страница 13: ... and press the knob to make your selection Select FREQ for the FUNCTION and 20KHZ 10000HZ 1000HZ or 2000CPM for the RANGE Disconnect all input wires from the devices to be calibrated and connect the PIECAL 820 to the input of the device matching polarity The green HZ SYNC LED pulses in synch with the output pulses and may be used to calibrate optical pickups The output signal is a zero crossing sq...

Страница 14: ...st misconnection Over voltage protection to 60 vrms rated for 30 seconds Red LED indicates OVERLOAD or out of range conditions Display High contrast graphic liquid crystal display with 0 315 8 0 mm high digits LED backlighting Read mA Ranges and Resolution 0 00 to 24 00 mA or 25 0 to 125 0 of 4 20 mA Accuracy 0 03 of Full Scale Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nom...

Страница 15: ...ance 1 Megohms Open TC Threshold Pulse 10K Ohms 5 μamp pulse for 300 milliseconds nominal RTD OHMS and Continuity Read Resistance Ranges 0 0 to 401 0 0 to 4010 Ohms Accuracy 0 03 of Full Scale 0 075 Ohms Excitation Current 1 0 mA to 401 Ohms 0 6 mA to 4010 Ohms nominal Continuity 0 0 to 401 0 Ohms Beeps from 0 0 to 100 0 Ohms RTD and OHMS Source Accuracy From 1 to 10 2 mA External Excitation Curre...

Страница 16: ...es C Range C Degrees F Range F J 200 0 to 150 0 1 2 328 0 to 238 0 2 0 150 0 to 50 0 0 7 238 0 to 58 0 1 3 50 0 to 100 0 0 5 58 0 to 212 0 0 9 100 0 to 1200 0 0 4 212 0 to 2192 0 0 8 K 230 0 to 150 0 2 6 382 0 to 238 0 4 7 150 0 to 0 0 1 0 238 0 to 32 0 1 8 0 0 to 1100 0 0 6 32 0 to 2012 0 1 1 1100 0 to 1371 1 0 7 2012 0 to 2500 0 1 2 T 260 0 to 230 0 6 1 436 0 to 382 0 11 0 230 0 to 150 0 2 2 382...

Страница 17: ... 0 to 1500 0 1 3 752 0 to 2732 0 2 4 1500 0 to 2000 0 2 3 2732 0 to 3632 0 3 0 2000 0 to 2320 0 2 6 3682 0 to 4208 0 4 6 D 1 1 to 200 0 1 8 30 0 to 392 0 3 2 200 0 to 1350 0 1 4 392 0 to 2462 0 2 6 1350 0 to 2000 0 1 9 1742 0 to 3632 0 3 4 2000 0 to 2320 0 2 6 3632 0 to 4208 0 4 7 P 200 0 to 0 0 2 1 328 0 to 32 0 3 8 0 0 to 150 0 0 8 32 0 to 302 0 1 5 150 0 to 1100 0 0 6 302 0 to 2012 0 1 1 1100 0...

Страница 18: ...y Our equipment is warranted against defective material and workmanship excluding batteries for a period of three years from the date of shipment Claims under warranty can be made by returning the equipment prepaid to our factory The equipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No ...

Страница 19: ...ated on equipment traceable to NIST and includes a of Calibration Test Data is available for an additional charge Practical Instrument Electronics recommends a calibration interval of one year Contact your local representative for recalibration and repair services Shelby Jones Co 8800 W Chester Pike Upper Darby PA 19082 610 446 6600 or Toll Free 800 346 4620 www Instrumentation com ...

Отзывы: