Configuration

Configure the Calibrator

Press and release the ON/OFF Key 

to turn 

the calibrator on then press and release the 
MENU Key 

to enable the configuration menu. 

Turn the EZ-Dial™ Knob to scroll through the list 
of configuration options. Press the EZ-Dial™ 
Knob to change configuration items.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

Default Settings Shown 

AUTO  OFF

 - pressing the knob will toggle 

between ON and OFF. If AUTO OFF is ON, the 
unit will turn off after 30 minutes of inactivity to 
save battery life. If AUTO OFF is OFF the unit 
will stay on until the ON/OFF key is pressed. Set 
If you are using the Totalizer functions for over 
30 minutes you should set AUTO OFF to OFF.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

X1/X10

 - pressing the knob will toggle between 

X1 and X10 attenuation while reading frequency. 
Choose X1 when you are measuring signals 
with amplitudes between 0.1 and 12 V peak and 
X10 when measuring between signals with 
amplitudes between 12 and 120 V peak.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

>AUTO OFF

 ON

 X1/X10 

 X10

 0 XING/BASED 

 BASED

 EZ-CHECK  

 ON

 SINE/SQ 

 SQ

 BASIC CONFIGURATION 

0 XING/BASED

 - pressing the knob will toggle 

between zero crossing and zero based output 
waveforms. Select zero crossing when the 
instruments you are calibrating require a 
waveform that has both positive and negative 
peaks that are centered around zero. Select zero 
based when the instruments you are calibrating 
require only positive signals.
Note: To simulate negative only signals select 
zero based and connect the red lead to the 
ground and the black lead to the input connection 
of the instrument being calibrated.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

EZ-CHECK

 - pressing the knob will toggle 

between ON and OFF. Select ON to have the   
highest and lowest measured frequencies 
recalled by moving the EZ-Check switch on the 
side of the calibrator to HI or LO. Select OFF and 
the currently measured freqency will be 
displayed regardless of the position of the 
EZ-Check switch.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

SINE/SQ

 - pressing the knob will toggle between 

SINE and SQ. Select SINE when the instruments 

you are calibrating require a sine wave and SQ 

when the instruments you are calibrating require 

a square wave or pulse input.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

BASIC  CONFIGURATION

 - pressing the knob 

will restore the 541 to the basic configuration 
with factory defaults as shown in the menu 
diagram. Output values stored in the EZ-Check 
HI and LO positions are not changed.
Press and release the MENU or STORE/CLEAR Key 
to save your selection.

Note:

 All settings are remembered even with the 

power off. Removing the batteries resets the 
configuration to factory defaults (Basic 
Configuration) and resets the EZ-Check HI and 
LO outputs to factory values.

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Содержание PIECAL 541

Страница 1: ...ptical pickups with the GATE TIME LED that is synchronized with the output pulses Match your instrument signal levels Choose zero based or zero crossing square or sine waves output from 0 1 V to 12 V peak to peak Measure from 0 1 V to 120 V peak with X1 or X10 attenuation Turn the dial to adjust the output amplitude to the level required Read signals from flow meters Measure flow meter vibration p...

Страница 2: ...ero Based Zero Crossing EZ Check HI LO Readings SINE SQUARE Wave output and reset to BASIC CONFIGURATION t READ GATE TIME Key Press and release READ GATE TIME key to step through the read ranges Read CPH 1 20000 Read CPM 0 1 2000 0 Read HZ 0 01 200 00 Read HZ 0 1 2000 0 Read KHZ 0 001 20 000 Totalizer Press and hold the READ GATE TIME key to enable time adjustment for Source Read Totalize mode Tur...

Страница 3: ...aks that are centered around zero Select zero based when the instruments you are calibrating require only positive signals Note To simulate negative only signals select zero based and connect the red lead to the ground and the black lead to the input connection of the instrument being calibrated Press and release the MENU or STORE CLEAR Key to save your selection EZ CHECK pressing the knob will to...

Страница 4: ...y Instruments SOURCE CPH 1 20000 20000 HI Storing EZ Check Outputs STORING HI and LO EZ Check Outputs 1 Store your high SPAN output frequency by moving the EZ Check switch q to the HI position and turn the EZ Dial knob u until the desired frequency is on the display Press the STORE CLEAR Key and STORED appears on the display 2 Store your low ZERO output frequency by moving the EZ Check switch q to...

Страница 5: ... 541 to the input of the device to be calibrated 5 Press the EZ Step push button q like a stop watch to start outputting pulses The word RUN will appear on the display and the 541 display increments with each pulse while the totalizer is running When the selected number of seconds has elapsed the word STOP appears on the display and the total number of pulses sourced during the time appears on the...

Страница 6: ... adjust the totalize time from 1 to 100 minutes Press and release the READ GATE TIME or STORE CLEAR button to save your selection 3 Connect the 541 to the output of the device to be calibrated 4 Turn the EZ Dial knob u to adjust the trigger level of the 541 The LED will light when the 541 locks onto the input signal indicating that it is properly detecting the signal 5 Press the EZ Step push butto...

Страница 7: ...Rejection 50 60 Hz 50 dB 50 60 Hz 90 dB Read Frequency Input Impedance 1 Megohms 100pF Signal Attenuation Adjustment Range Adjustable trigger level with X1 0 1 V peak to 12 0 V peak and X10 1 V Peak to 120 V peak Note on reading slow sine waves High signal noise and low slew rate Volts per second will affect reading uncertainty Source Frequency Output Impedance 25 Ohms Source Current 6 mA at 12 V ...

Страница 8: ...warranty can be made by returning the equipment prepaid to our factory The equipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No responsibility is accepted for damage loss or other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electro...

Отзывы: