PIE 820PM Скачать руководство пользователя страница 7

Page 7

Practical Instrument Electronics

Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

To change the Automatic Stepping settings

Double click the 

e

 DIAL KNOB at any time the unit is on and the menu will appear for 15 seconds.

Turn the 

e

 DIAL KNOB to move through down 

to the third (FEATURES) menu. Press the 

e

 

DIAL KNOB to toggle between OFF and ON or 
to change the STEPS and the STEP TIME 
settings. These settings are remembered even 
with the power off. 

EXIT MENU

 - exits this menu immediately and saves any changes. Menu will automatically exit after 

15 seconds of inactivity.

STEPS

 - pressing the knob will cycle through 2, 3, 5 and 11 then reverse direction. The endpoints of 

the steps are based on the values stored in the 

HI

 and 

LO

 EZ-CHECK outputs. 

2 steps

 will automatically switch between the values stored in the HI & LO EZ-CHECK (0 & 

100%). 

 

3 steps

 between the HI, Midpoint and LO EZ-CHECK (0, 50 & 100%). 

5 steps

 between the HI and LO EZ-CHECK in 25% increments (0, 25, 50, 75 & 100%).

11 steps

 between the HI and LO EZ-CHECK in 10% increments (0, 10, 20...80, 90 &100%).

RAMP

 continuously between the HI and LO EZ-CHECK.

STEP TIME

 - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds. 

To start the Automatic Stepping

Start automatic stepping  or ramping by placing the EZ-CHECK Switch into the HI or LO position 
then press and hold the 

e

 DIAL KNOB for 6 seconds  (the word STORE will appear on the display 

after 3 seconds and continue to press the DIAL KNOB) until the word STEPPING appears on the 
display.  The word STEPPING will appear on the display anytime the selected automatic function is 
running. Stop the stepping by again pressing and holding the 

e

 DIAL KNOB for 3 seconds. 

Automatic Stepping

Storing EZ-CHECK Outputs

STORING HI and LO EZ-CHECK Outputs

Choose this function to provide a simulated thermocouple signal into controllers, temperature 
transmitters, indicators or any other input device that measure thermocouple sensors.

1)  Store your high (SPAN) output temperature by moving the EZ-CHECK switch to the 

HI

 position 

and turn the 

e

 EZ-Dial knob until the desired temperature is on the display. Press and hold the 

EZ-Dial knob until 

STORED

 appears to store the value. Release the EZ-Dial knob.

2)  Store your low (ZERO) output temperature by moving the EZ-CHECK switch to the 

LO

 position 

and turn the 

e

 EZ-Dial knob until the desired temperature is on the display. Press and hold the 

EZ-Dial knob until 

STORED

 appears to store the value. Release the EZ-Dial knob.

3) Instantly output your SPAN and ZERO temperature outputs by moving the EZ-CHECK switch 

between HI and LO. You may also select any third temperature output (such as mid-range) using 
the SET position on the EZ-CHECK switch.

           

FEATURES

> EXIT (3/3)   

      

   AUTO OFF  

        ON 

OFF

   BACKLIGHT   

   ON 

OFF

   STEPS/RAMP         2

 

3 5 11 RAMP

   STEP/RAMP TIME   5 

6 7 8 9 10 15 20 25 30 60

Содержание 820PM

Страница 1: ...piecal com www piecal com PIE 820PM Panel Mounted Multifunction Process Calibrator mA V TC RTD Hz Operating Instructions PIE 820PM Panel Mounted Multifunction Process Calibrator mA V TC RTD Hz Operati...

Страница 2: ...Simulate 2 Wire Transmitters 8 9 Read mA Power Measure Transmitters 10 11 Voltage Millivolt Source V mV Read V mV 12 13 Thermocouple Source T C Read T C Sensors 14 15 Resistance Source Resistance Rea...

Страница 3: ...3 outputs Plus automatic step ramp Easily set any value quickly with the adjustable DIAL plus store any three temperatures for instant recall with the EZ CHECK switch Choose between 2 3 5 11 steps and...

Страница 4: ...ockwise to decrease the output in one least significant digit step at a time Push down and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI...

Страница 5: ...late 2 Wire Transmitters Read mA Power Measure Transmitters Source Volts Millivolts Read Volts Millivolts EXIT 1 2 FUNCTION FREQ RANGE 20KHZ 10000HZ 1000HZ 2000CPM Source Read Frequency EXIT 1 2 FUNCT...

Страница 6: ...or in areas with enough ambient light to read the display FEATURES EXIT 2 2 AUTO OFF ON BACKLIGHT ON STEPS RAMP 3 STEP RAMP TIME 5 MAIN EXIT 1 2 FUNCTION mA MODE SOURCE UNITS mA HART 250 ON STEPS RAMP...

Страница 7: ...LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the word STEPPING appears on the disp...

Страница 8: ...jack 5 to the plus input and black lead from jack 6 to the minus input Open loops and signals above the maximum scale are limited by protection circuitry with ERROR or OVER RANGE flashed on the displ...

Страница 9: ...lead from jack 6 to the minus input Open loops and signals above the maximum scale are limited by protection circuitry with ERROR or OVER RANGE flashed on the display and the red OVERLOAD LED lit Inst...

Страница 10: ...nput lead from jack 6 to the more negative point Signals below 0 mA or open circuits are indicated by 0 00 mA 25 0 on the display Signals above 24 mA are current limited by protection circuitry with O...

Страница 11: ...Connect the red source lead of the PIE 820PM from jack 5 to the plus input of the device and the black source lead from jack 6 to the minus The PIE 820PM supplies a nominal 24 volts DC at 24 mA to th...

Страница 12: ...d making sure to check polarity Connect the Red lead from jack 1 to the plus input and black lead from jack 2 to the minus input Instantly output your SPAN and ZERO output settings by moving the EZ CH...

Страница 13: ...the black lead to jack 2 When measuring up to 10 V or 60 V connect the red lead to jack 5 and the black lead to jack 6 Signals above the maximum scale are limited by protection circuitry with OVER RAN...

Страница 14: ...the inputs of the device being calibrated using the proper type of thermocouple wire via the miniature thermocouple socket Instantly output your SPAN and ZERO output settings by moving the EZ CHECK sw...

Страница 15: ...rmocouple wire via the miniature thermocouple socket If no sensor is connected a wire is broken or the sensor is burned out OPEN TC will appear on the display Signals above the maximum scale are limit...

Страница 16: ...nto jack 1 and the black lead into jack 2 For 3 wire resistance add a black lead plugged into jack 3 and for 4 wire resistance add a red lead plugged into jack 4 Instantly output your SPAN and ZERO ou...

Страница 17: ...0PM to the inputs of the device using 2 3 or 4 wires If continuity is selected resistance is measured up to 400 0 The beeper will sound and appears on the display when resistances 100 0 are measured S...

Страница 18: ...ice using 2 3 or 4 wires For 2 Wire RTDs connect the red lead into jack 1 and the black lead into jack 2 For 3 wire RTDs add a black lead plugged into jack 3 and for 4 wire RTDs add a red lead plugged...

Страница 19: ...PM to the inputs of the device using 2 3 or 4 wires For 2 Wire RTDs connect the read lead into jack 1 and the black lead into jack 2 For 3 wire RTDs add a black lead plugged into jack 3 and for 4 wire...

Страница 20: ...e green HZ SYNC LED pulses in synch with the output pulses and may be used to calibrate optical pickups The output signal is a zero crossing square wave with a fixed amplitude of 6 V peak to peak from...

Страница 21: ...d black lead from jack 2 to the minus input The green HZ SYNC LED pulses in synch with the input frequency Signals above the maximum scale are limited by protection circuitry with OVER RANGE on the di...

Страница 22: ...s restricted to that given under our warranty No responsibility is accepted for damage loss or other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument...

Страница 23: ...rt No 020 0820 1 Red 1 Black Lead with Banana Plug Alligator Clips 2 Red 2 Black Leads with Banana Plugs Spade Lugs OPTIONAL T C Wire Kit 1 for Types J K T E Part No 020 0202 T C Wire Kit 2 for Types...

Страница 24: ...ce 1 Ohm Short Circuit Duration Infinite Thermocouple Source Accuracy 0 03 of Full Scale Note Full Scale is 80 00 mV Cold Junction Compensation 0 1 C Thermistor traceable to NIST for 11 years Output I...

Страница 25: ...or pulsed intermittent Pulsed Excitation Current Compatibility DC to 0 01 second pulse width 0 025 mV mA Excitation Current Frequency Source Ranges 1 to 2000 CPM 0 01 to 999 99 Hz 0 1 to 9999 9 Hz 0...

Страница 26: ...5 122 0 to 572 0 1 0 300 0 to 400 0 0 4 572 0 to 752 0 0 7 E 240 0 to 150 0 2 5 400 0 to 238 0 4 5 Chromel Constantan Jacket Purple Red Purple 150 0 to 50 0 0 7 238 0 to 58 0 1 1 50 0 to 150 0 0 4 58...

Страница 27: ...0 2 6 1350 0 to 2000 0 1 9 1742 0 to 3632 0 3 4 2000 0 to 2320 0 2 6 3632 0 to 4208 0 4 7 D W3 1 1 to 400 0 2 5 30 0 to 752 0 4 5 W3 Re W25 Re Jacket White Red White Yellow 400 0 to 1500 0 1 3 752 0 t...

Страница 28: ...00 0 0 9 1 0 1 2 Pt 100 Ohm Old JIS 1981 1 3916 200 0 to 170 0 170 0 to 480 0 480 0 to 648 9 0 5 0 6 0 7 328 0 to 338 0 338 0 to 896 0 896 0 to 1200 0 0 9 1 0 1 2 Pt 100 Ohm US Lab 1 3926 200 0 to 180...

Страница 29: ...ractical Instrument Electronics Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com 0 565 0 995 4 375 7 755 8 185 8 85 0 196 0 35 4 40 4 554 4 75 0 125 Dia x6 Bezel Cutout Panel Cutout D...

Страница 30: ...Page 30 Practical Instrument Electronics Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com 820PM mounted in a panel...

Страница 31: ...Page 31 Practical Instrument Electronics Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Mounted in Panel Rear view of 820PM mounted in a panel with AC Adapter...

Страница 32: ...Page 32 Practical Instrument Electronics Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com...

Отзывы: