4
How to use this Manual
Remark:
The LDP-CWL 06-20 and LDP-CWL 12-20 described in this manual
are base plate cooled laser diode drivers. Improper cooling may cause an
internal overtemperature shutdown (max. 80 °C).
Base plate cooling:
Depending on the final application and operation mode, this
unit may stay non-cooled or must be assembled onto a heat sink. Please refer to
section “Working Areas and Power Dissipation” for more details about the thermal
power losses during operation.
Before powering on your unit, read this manual thoroughly and make sure you
understood everything!
Please pay attention to all safety warnings.
If you have any doubt or suggestion, please do not hesitate to contact us!
Dos and Don’ts
Keep the cable connection between power supply and laser diode driver
and the connection between driver and laser diode as short as possible to
minimize the stray inductance.
Please set the laser diode driver only in the scheduled working area and only while
the power supply is turned off. Therefore please have a look at the section
“Description of the Connectors of the LDP-CWL 06-20 / LDP-CWL 12-20”.
Never disconnect or connect the load while the driver is operational, as it destroys
the driver and/or the connected load. Make sure that the control ground is
connected to the supply ground. Any unexpected current flow through the control
board may damage the driver.
Mount the driver on an appropriate heat sink.
Please be aware that there might be hot surfaces, be careful not to touch
them!