background image

Voltage v frequency rating curve

10 kHz

0

100

200

300

400

500

600

700

100 kHz

1 MHz

10 MHz

100 MHz

1 GHz

WARNING

This probe does not carry a measurement category rating. To prevent 

electric shock, do not connect to a mains (line) voltage or a derived 

voltage that can carry the overvoltage transients that may be present.

Measurement categories are defined in IEC 61010-031 as follows:

No Measurement Category (not in CAT II, III or IV)

Definition For measurements performed on circuits not directly connected 

to a mains supply.

Grounding

WARNING

Never connect the ground input to, or allow it to touch, any electrical 

potential other than ground. To prevent personal injury or death, use a 

voltmeter to check that there is no significant AC or DC voltage between 

the probe ground and the point to which you intend to connect it.

CAUTION

Applying a voltage to the ground input is likely to cause permanent 

damage to the probe or other connected equipment.

It is good practice to connect the probe output to the measurement instrument 

and the ground lead to earth ground before connecting the probe to the circuit 

under test. Disconnect the probe input and the probe ground lead from the 

circuit under test before disconnecting the probe from the measurement 

instrument.

Содержание TA386

Страница 1: ...User s Guide TA386 X1 X10 Max 600 Vp X 1 X 1 0 TA375 100 MHz oscilloscope probe TA386 200 MHz oscilloscope probe...

Страница 2: ...es including damages for loss of profits loss of business loss of use or data interruption of business and the like even if Pico has been advised of the possibility of such damages arising from any de...

Страница 3: ...input ranges The table and frequency derating plot below indicate the full scale measurement range and overvoltage protection range for these probes The full scale measurement ranges are the maximum v...

Страница 4: ...ng WARNING Never connect the ground input to or allow it to touch any electrical potential other than ground To prevent personal injury or death use a voltmeter to check that there is no significant A...

Страница 5: ...There may be a charge for these services unless covered by the Pico two year warranty WARNING To prevent injury or death do not use the product if it appears to be damaged in any way and stop use imme...

Страница 6: ...ntil you see a flat top square wave on the display CORRECT INCORRECT INCORRECT Accessories and features The probe is provided with several accessories designed to make probing and measurement simpler...

Страница 7: ...Ground lead Use the alligator clip to attach the probe to a ground reference Hook tip Spring loaded for gripping wires and test points Adjustment tool Use to adjust the probe compensation BNC connect...

Страница 8: ...0 MHz 200 MHz Rise time calculated 35 ns 3 5 ns 35 ns 1 75 ns Input resistance 1M 10 M 2 1M 10 M 2 Input capacitance 57 pF CS 15 pF 57 pF CS 15 pF Max working voltage 42 4 V pk 600 V pk pk 42 4 V pk 6...

Отзывы: