8
Product information
2
Product information
2.1
Specifications
PicoLog 1012
PicoLog 1216
Maximum sampling rate
continuous streaming
block mode
[1]
100 kS/s single-channel
1 MS/s single-channel
Buffer size
Analog inputs
analog bandwidth
(-3 dB)
input type
input voltage range
8k sample internal buffer shared by all
channels;
1 million sample limit using SDK
12
[2]
16
[2]
DC to 70 kHz
Single-ended, unipolar
0 to +2.5 V
linearity
(at 25 °C)
1 LSB
resolution
10 bits
12 bits
accuracy
1%
0.5%
overload protection
±30 V to ground
input coupling
DC
input impedance
1 MΩ
Digital outputs (D0...D3)
2
4
[2]
Digital outputs (
PWM
)
None
1
period
duty cycle
100 µs to 1800 µs
adjustable from 0% to 100% in 1% steps
Digital outputs (all)
logic low voltage
logic high voltage
current limiting
Power output for sensors
Ground fault current protection
I/O connector
Environmental operating
conditions
for quoted accuracy
general operation
relative humidity
100 mV (typical)
3.3 V
1 kΩ resistors in series with outputs
2.5 V @ 10 mA, current-limited
0.9 A thermal self-resetting fuse
25-way D female
20°C to 50°C for quoted accuracy
0°C to 70°C overall
5% to 80% RH
Compliance
CE (EMC) class A emissions & immunity
FCC emissions
PC connection
USB
Dimensions
Weight
45 mm x 100 mm x 140 mm
(1.77" x 3.94" x 5.51")
< 200 g (7.05 oz)
Notes:
1. Block mode is supported by PicoScope and the PicoLog 1000 Series API. The
PicoLog software will allow a Fast Block capture with one PicoLog 1000 Series
device connected to the PC.
2. Except in
USB ADC-11 compatibility mode
(see
I/O connections table
)
pl1000.en r3
Copyright © 2013–2016 Pico Technology Ltd. All rights reserved.