Reference Values
The blood glucose readings deliver plasma equivalent results.
Time of day
Normal plasma glucose range for
people without diabetes (mg/dL)
Fasting and before meal
< 100 mg/dL (5.6 mmol/L)
2 hours after meals
< 140 mg/dL (7.8 mmol/L)
Source: American Diabetes Association. Standards of Medical Care in Diabetes- 2018 Jan; 41(Supplement 1): S1-S2.
Please consult your doctor to determine a target range that works best for you.
SYSTEM TROUBLESHOOTING
If you follow the recommended action but the problem persists, please call Consumers’ toll-free number for Italy
800 900 080 and for France 01.55.93.26.48.
Error Messages
MESSAGE
WHAT IT MEANS
WHAT TO DO
E-1
Appears when the batteries are too
low.
Replace the batteries immediately.
E-2
Appears when a used test strip is
inserted.
Repeat with a new test strip.
E-3
Appears when test strip is removed
while counting down, or insufficient
blood volume.
Review the instructions and repeat test with a
new strip.
If the problem persists, please contact the
Consumers’ toll-free number for Italy 800 900
080 and for France 01.55.93.26.48. for help.
E-4
Appears when ambient temperature
is above or below system operation
range.
System operation range is 8°C to 45°C (46.4°F
to 113°F). Repeat the test after the meter and
test strip are in the above temperature range.
E-5
E-6
Problem with the meter.
Repeat the test with a new test strip. If the
meter still does not work, please contact the
Consumers’ toll-free number for Italy 800
900 080 and for France 01.55.93.26.48. for
assistance.