15
• Fit the filter cover cap back on.
WARNING! Never leave the filter cover cap and/or filters unattended: These are small parts posing the risk of
suffocation if swallowed by children.
TROUBLESHOOTING
PROBLEM
CAUSE
SOLUTION
Scarce nebulisation
Nebuliser cup clogged
Remove and disconnect all nebu-
liser cup components and clean
them according to the instruc-
tions provided in this instruction
manual.
Scarce nebulisation
Nebuliser cup clogged
If the problem remains after clean-
ing, replace the nebuliser cup.
No nebulisation
Components of the nebuliser cup
assembled incorrectly
Disassemble and reassemble the
nebuliser cup. If the problem re-
mains, replace the nebuliser cup.
No nebulisation
The compressor does not work
Check whether the AC/DC adapter
plug is disconnected from the
mains socket, check that the ON/
OFF button is ON, that the mains
socket is operational, and that the
AC/DC adapter power supply ca-
ble pin is connected to the device
socket (N).
Slow nebulisation
Medicine too thick
Dilute the medicine as specified
by the pharmaceutical company.
TECHNICAL SPECIFICATIONS
External power supply:
100-240 VAC 50/60Hz 0.6A 12V 2A
02010298000000 (EU plug)
02010299000000 (UK plug)
Service life in storage conditions:
5 years
Service life of accessories:
1 year
Frequency:
50/60
Hz
Maximum flow:
12 ± 2l/min
Operational flow:
5 ± 1l/min
Max pressure:
2 ± 0.4 bar
Operational pressure:
0.6 ± 0.18 bar
Conditions of use:
20’ ON – 40’ OFF at 40°C
NEB RATE:
> 0.30 ml/min
MMAD:
< 3 μm
Weight of the Mister 8 [without accessories]: ~800g
Nebuliser cup capacity:
Max 10ml; Min 2ml
Breathable fraction:
> 75%