ENGLISH
GB
15
Radiated/Conducted Emissions
CISPR11
Class [B]
The Miss APPY aerosol therapy
unit is suitable for use in all
environments, including do-
mestic ones and those directly
connected to the public power
supply network that supplies
buildings used for domestic pur-
poses.
Harmonic Emissions
IEC/EN 61000-3-2
Class [A]
Voltage Fluctuations/Flicker
Emissions
IEC/EN 61000-3-3
[Satisfi es]
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Miss APPY aerosol therapy unit is intended for use in the electromagnetic environment specifi ed
below. The client or user of the Miss APPY aerosol therapy unit shall make sure that the device is
used in this environment
Emissions test
Compliance
Electromagnetic
Environment - Guidance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
±6kV on contact
±8kV in the air
Floors should be wood, concre-
te or ceramic tile. If fl oors are
covered with synthetic material,
the relative humidity should not
exceed 30%.
Electrical fast transient/burst
IEC/EN 61000-4-4
2 kV power supply
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC/EN 61000-4-5
1 kV differential mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short interruptions
and voltage variations
IEC/EN 61000-4-11
<5%U
t
UT for 0.5 cycles
40% U
t
for 5 cycles
70% U
t
for 25 cycles
<5% U
t
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Miss APPY aerosol therapy
unit requires continued opera-
tion, it is recommended that the
unit is powered from an uninter-
ruptible power supply.
Magnetic fi eld
IEC/EN 61000-4-8
3A/m
The magnetic fi eld should be
that of a typical commercial or
hospital environment.
Immunity to conducted distur-
bances
IEC/EN 61000-4-6
3V/m 80MHz to 2.5GHz (for non-
life-supporting equipment)
Irradiated immunity
IEC/EN61000-4-3
3V/m 80MHz to 2.5GHz (for non-
life-supporting equipment)
NoTe: Ut is the mains voltage
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Miss APPY aerosol therapy unit is intended for use in the electromagnetic environment specifi ed
below. The client or user of the Miss APPY aerosol therapy unit shall make sure that the device is used
in this environment
Emissions test
Compliance
Electromagnetic
Environment - Guidance
Radiated/Conducted Emissions
CISPR11
Group 1
The Miss APPY aerosol therapy
unit only uses RF energy for its
internal function. Therefore its
RF emissions are very low and
do not cause any interference in
nearby electronic equipment.