5 Installation
26
Version: 1.1.0
PZ268D
P-616 NanoCube® XYZ Nanopositioner
5.6
Connecting the P-616 to the Electronics
INFORMATION
Systems consisting of a P-616 and electronics are calibrated at the factory to achieve optimum
performance.
Pay attention to the assignment of the axes to the channels that are specified on the
calibration label of the electronics.
Requirements
You have read and understood the general notes on installation (p. 15).
You have installed suitable electronics (p. 11).
You have read and understood the user manual of the electronics.
The electronics are switched off, i.e.,
not
connected to the power supply.
Tools and accessories
When connecting to electronics with LEMO sockets: P-895.3DLC (p. 10) adapter cable
Connecting the P-616 to the E-727 controller
1.
Plug the connector of the P-616 into the corresponding socket of the controller (see
user manual of the controller).
2.
Use the integrated screws to secure the connection against accidental disconnection.
Connect the P-616 to the E-50x modules or E-663.00 piezo amplifier with the P-
895.3DLC adapter cable
1.
Connect the Sub-D 25W3 (m) connector of the P-616 to the Sub-D 25W3 (f) connection
of the adapter cable.
2.
Connect the piezo connections of the adapter cable to the piezo connection of the E-
503.00 or E-663.00 piezo amplifier module as follows.
−
PZT1
to
PZT
for channel 1 (
CH1
)
−
PZT2
to
PZT
for channel 2 (
CH2
)
−
PZT3
to
PZT
for channel 3 (
CH3
)
3.
When connecting to the E-50x modules: Connect the sensor connections of the adapter
cable to the E-509.C3A control module as follows:
−
T1
to
T
for channel 1
(SERVO 1)
−
P1
to
P
for channel 1
(SERVO 1)
−
T2
to
T
for channel 2 (
SERVO 2
)
−
P2
to
P
for channel 2 (
SERVO 2
)
−
T3
to
T
for channel 3 (
SERVO 3
)