6.6
Connecting the L-505.011200
NOTICE
Damage due to mechanical stress on the cable!
The cable will break from excessive or repeated bending. A broken cable will lead to failure
of the L-505.011200 or damage to the L-505.011200 or the electronics.
►
Install the L-505.011200 so that the cable has a bending radius of
≥
35 mm.
►
If necessary: Use cables that are approved for recurring mechanical loading (e.g., drag
chain cables).
Tools and Accessories
■
If necessary: Suitable screwdriver for the locking screws of the connectors.
Requirements
✓
You have read and understood the
general safety instructions (p. 8)
.
✓
You have read and understood the user manual for the electronics used.
✓
You have installed the electronics properly.
✓
The electronics are switched off.
NOTICE
Damage due to incorrect connection of the L-505.011200!
Connecting unsuitable electronics or a wrong cable can damage the L-505.011200 or the
electronics.
►
Connect the L-505.011200 to
suitable electronics (p. 10)
from PI miCos only.
►
Use cables from PI miCos only to connect the L-505.011200 to the electronics.
Connecting the L-505.011200
1. If necessary: Remove the ESD protective caps from the connections of the L-505.011200.
2. Connect the drive plug of the L-505.011200 to the drive socket on the electronics.
3. Secure the connector against unintentional removal.
6 INSTALLATION
L505M0004EN
‒
11/26/2018
23