Connecting the Mechanics to the Protective Earth Conductor
INFORMATION
Observe the applicable standards for mounting the protective earth conductors.
F-712.MA2: In a double-sided system, connect both stacks to the protective earth
conductor.
The NanoCube® nanopositioner has an M4 hole for mounting the protective earth conductor.
This hole is marked with the symbol for the protective earth conductor.
Figure 15: Position of the protective earth connection
Requirements
You have read and understood the safety precautions (p.7).
The NanoCube® nanopositioner and the M-122K025 linear stages are not connected to
the controller.
Tools and accessories
Suitable protective earth conductors: Cross-
sectional area of the cable ≥0.75
mm
2
Supplied M4 protective earth screw set for connecting the protective earth conductor
Suitable screwdriver