NOTICE
Operating voltage too high or incorrectly connected!
Operating voltages that are too high or incorrectly connected can cause damage to the M-
122K025 linear stage.
Do
not
exceed the operating voltage range for which the M-122K025 linear stage is
specified.
Only operate the M-122K025 linear stage when the operating voltage is properly
connected, see “Pin Assignment” (p. 45).
NOTICE
Damage or considerable wear due to high accelerations!
High accelerations can cause damage to or considerable wear on the mechanical system.
Stop the motion immediately if a controller malfunction occurs.
Ensure that the end of the travel range is approached at low velocity.
Determine the maximum velocity for your application.
NOTICE
Failure of the M-122K025 linear stage!
Frequent motions over a limited travel range can cause the lubricant to be unevenly
distributed in the drive train. The unevenly distribution of the lubricant can lead to failure of
the M-122K025 linear stage.
A maintenance run over the entire travel range will evenly spread the lubricant again.
Carry out a maintenance run over the entire travel range at regular intervals. The more
often motions are carried out over a limited travel range, the shorter the time between the
maintenance runs has to be.