0x1F0007A4 PiezoMike Enter Window (Phys. Unit)
Used in: Motion in closed-loop operation (27)
Description:
Entrance window of the target position for motion in closed-loop
operation.
Commanded motion is done until the entrance window for the tar-
get position has been reached. The target position is then consid-
ered as reached and the on-target signal is output, if necessary, after
a delay time (parameter
).
Data type
FLOAT
Command level
1
Item type
Axis
Source of data
Positioner database
0x1F0007A5 PiezoMike Exit Window (Phys. Unit)
Used in: Motion in closed-loop operation (27)
Description:
Exit window of the target position for motion in closed-loop opera-
tion.
Defines a tolerance range around the entrance window for the tar-
get position (parameter
) to maintain the on-tar-
get signal after motion. The on-target signal is no longer output and
an attempt is made to approach the target position again only when
the current position is so far away from the target position that it
lies outside the exit window.
Data type
FLOAT
Command level
1
Item type
Axis
Source of data
Positioner database
10 PARAMETER REFERENCE
PZ285EN
‒
1/30/2019
207