6 Start-Up
E-836.1G Piezo Amplifier
PZ250E
Version: 1.0.1
19
6.3 Executing Motions
INFORMATION
The control signal for the output piezo voltage is the sum of the following signals:
Control input voltage at the
Control In
BNC socket
Offset that is set with the
DC Offset
potentiometer
The control signal should always be within the range of –2 to +12 V. The range can
be expanded to –3 to +13 V. However, this can shorten the lifetime of the piezo
actuator in the stage (p. 35).
Prerequisite
You have read and understood the user manual of the stage.
You have mounted the stage according to the description in the corresponding
user manual.
You have connected the stage to the E-836.1G (p. 15).
When the stage is operated via a control input voltage: You have connected a
signal source to the E-836.1G (p. 15).
You have switched on the E-836.1G (p. 18).
Executing motions
Change the control signal within the range of -2 V to +12 V:
−
Change the control input voltage at the
Control In
BNC socket.
−
Turn the
DC Offset
potentiometer.
The stage generates corresponding motions.