PI E-614.2BS Скачать руководство пользователя страница 17

Calibration 

 

www.pi.ws 

E-651 / E-614 Controller / Amplifier PZ129E Release 6.1.0  Page 15 

channel accessible through the front panel, p. 8; E-614: R132 
for channel 1, R133 for channel 2). This potentiometer shifts the 
output of the sensor processing circuitry and hence the values 
on the sensor monitor output (E-651: front panel; E-614: MON-
DMS1 and MON-DMS2 test points) and sensor-input lines of 
the servo-loop submodule. 

For the operating elements of the E-651 benchtop devices see 
p. 8. 

For the location of the potentiometers, connectors, switches 
and jumpers of the E-614 OEM board see the E614T0003 
technical note. 

To adjust the zero-point for a channel proceed as follows: 

Before powering up the system, make sure the actuator 
is mounted in the same way and with the same load as 
during normal operations in the application. In multi-
axis systems, make sure the actuators are always 
connected to the same controller channels.  

Make sure the analog control input (E-651: front panel; 
E-614: connector X20) is at 0 V. 

Set the channel to open-loop mode (E-651: SERVO 
OFF; E-614: EC/VC switch to VC) 

E-614 only: close jumper X7 (channel 2) and X10 
(channel 1) to disable any DC offset potentiometer. 

Power up the system. 

Deflect the actuator over its full range by sweeping the 
control input signal over the range of -5 V to +5 V.  

Hold the control input signal at -5 V. 

Adjust the ZERO potentiometer so that the sensor-
monitor signal is 0 V. 

Normally the zero-point adjustment can be finished at this point. 
Now you can switch the channel to closed-loop mode and 
deflect the actuator over the full range by applying a control 
input voltage of -5 to +5 V. The overflow LED should not light 
up. If the overflow LED does light, repeat steps 2 to 8. 

Failure of the zero-point re-adjustment (i.e. overflow LED still 
lights up) could indicate a serious problem. See "Calibration" for 
information on how to proceed. 

Содержание E-614.2BS

Страница 1: ...s Release 6 1 0 Date 2019 11 18 This document describes the following Product s E 651 1S Piezo amplifier servo controller for bending actuators 1 channel 0 to 60 V strain gauge sensor benchtop device...

Страница 2: ...ration of the electrical device into the overall system 1999 2019 Physik Instrumente PI GmbH Co KG Karlsruhe Germany The text photographs and drawings in this manual are protected by copyright With re...

Страница 3: ...e of Delivery 8 3 4 Product View 8 4 Operation 11 4 1 Getting Started 11 4 2 Operating Modes 12 4 2 1 Closed Loop Operation 12 4 2 2 Open Loop Operation 13 4 3 On Board Submodules 13 4 3 1 E 801 Submo...

Страница 4: ...ts 7 3 Ambient Conditions and Classifications 20 7 4 Pin Assignment 20 7 4 1 E 651 PZT SENSOR connector 20 7 4 2 C 501 15050H Power Supply 21 8 Cleaning 22 9 Customer Service 23 10 Old Equipment Dispo...

Страница 5: ...atic as well as dynamic applications 1 2 General Safety Instructions The E 651 E 614 is built according to state of the art technology and recognized safety standards Improper use can result in person...

Страница 6: ...ore charge precautions must also be taken when the E 614 OEM module is not powered After switching off the E 614 OEM module wait a minute before you touch parts of the E 614 OEM module to be sure that...

Страница 7: ...ll information given by the manufacturer to the user documentation for example supplements or Technical Notes If you pass the E 651 E 614 on to other users also turn over the user documentation as wel...

Страница 8: ...er Manual E 802 Servo Control Submodule PZ150E User Manual E 614 2BS OEM module E614T0003 Technical Note 2 2 Downloading Manuals 1 Open the website www pi ws 2 Search for the product number e g P 871...

Страница 9: ...0 and P 871 from PI The PICMA bending actuators require full differential voltage control with constant voltages of 0 V and 60 V and a variable voltage from 0 V to 60 V see Fig 1 Note that a control i...

Страница 10: ...nality as the E 651 2S 2 channel controller but on a printed circuit board 3 3 Scope of Delivery PZ129E User manual for E 651 E 614 this document E 651 benchtop devices only C 501 15050H Wide range 15...

Страница 11: ...ller SERVO ON OFF Switch for changing between open loop off and closed loop mode on CONTROL IN BNC connector for the analog control signal 5 V to 5 V SENSOR MONITOR BNC connector for reading out the s...

Страница 12: ...E 651 E 614 Controller Amplifier PZ129E Release 6 1 0 Page 10 Fig 3 E 651 rear panel details POWER ON OFF switch 14 to 16V DC INPUT Operating power input socket for the barrel connector from the C 501...

Страница 13: ...n page 14 4 1 Getting Started This section gives instructions for the E 651 benchtop devices Instructions for E 614 OEM modules can be found in the E614T0003 technical note Do not interchange controll...

Страница 14: ...monitor the actuator deflection with the SENSOR MONITOR signal see step 2 0 V to 10 V even when the system is in open loop mode If the output voltage corresponding to the control input signal is outs...

Страница 15: ...evaluation of one strain gauge sensor SGS It is designed as a plug in submodule to be used on different PCBs The submodule supplies the strain gauge with a DC voltage and generates a signal proportio...

Страница 16: ...on2 require special equipment and should only be done by qualified personnel and are required only in special circumstances See the user manuals for the sensor processing and servo control submodules...

Страница 17: ...make sure the actuators are always connected to the same controller channels 2 Make sure the analog control input E 651 front panel E 614 connector X20 is at 0 V 3 Set the channel to open loop mode E...

Страница 18: ...roller note that if you want to use the new actuator in regular operation a fully recalibration of the system is required and contact our customer service department p 23 Control input voltage source...

Страница 19: ...r panel and carefully remove the rear panel 3 Use the Torx 8 driver to unscrew the 2 screws in the blue frame on the front panel Carefully remove the front panel which also pulls the PCB out of the ho...

Страница 20: ...er type P I analog P I analog Sensor type SGS SGS Sensor bandwidth Low pass filter frequency 300 Hz 1 kHz 3 kHz selectable Low pass filter frequency 300 Hz 1 kHz 3 kHz selectable Amplifier Input volta...

Страница 21: ...ensor monitor socket BNC BNC Pin header Miscellaneous Operating temperature range 5 to 50 C 5 to 50 C 5 to 50 C Overheat protection Deactivation at 75 C Deactivation at 75 C Deactivation at 75 C Dimen...

Страница 22: ...nearly to 50 relative humidity at 40 C Operating temperature 5 to 50 C Storage temperature 0 C to 70 C Transport temperature 25 C to 85 C Overvoltage category II Protection class I Degree of pollution...

Страница 23: ...Technical Data www pi ws E 651 E 614 Controller Amplifier PZ129E Release 6 1 0 Page 21 7 4 2 C 501 15050H Power Supply 3 34A 15VCD output 100 240 VAC input 1 Center pin 15VDC 2 1 mm diameter 2 GND...

Страница 24: ...ning E 651 bench top units only 1 Disconnect the E 651 from the power source 2 Wait a minute to be sure that any residual voltage has dissipated 3 Clean the housing surfaces of the E 651 using mild de...

Страница 25: ...ng information Product and serial numbers of all products in the system Firmware version of the controller if applicable Version of the driver or the software if applicable Operating system on the PC...

Страница 26: ...our old equipment according to international national and local rules and regulations In order to fulfil its responsibility as the product manufacturer Physik Instrumente PI GmbH Co KG undertakes envi...

Отзывы: